PPCK+™ Operation and Maintenance Manual
©2001
DH Instruments, Inc
.
Page 58
OPERATION
PPCK+ uses self diagnostics to determine when to run the TPC configuration
routine in dynamic control mode. Therefore, it is not normally necessary to run
the configuration routine separately.
Running the Config routine may cause the set pressure to increase or decrease
by up to 1.4 MPa (200 psi).
The dynamic control mode volume determination routine may be run at any
pressure greater than 1 MPa (150 psi). Dynamic pressure control mode must
be active.
To run the volume determination routine,
first set a pressure above 1 MPa (150 psi).
Then press
[SPECIAL]
and select
<4Internal>
,
<1Config>
. The display is:
Selecting
<1Yes>
causes the volume determination routine to run. While the
routine is running,
<Running config…>
flashes on the bottom line of the
PPCK+ display. Completing the routine may take up to 2.5 minutes and cause
the pressure to change by 0.7 to 1.4 kPa (100 to 200 psi).
When the volume determination is complete
the display is:
Select
<1Yes>
to activate the new volume determination. Select
<2No>
to retain
the old volume determination and return to the main run screen.
The volume configuration routine cannot be run under 1 MPa (1 000 psi). An error
message is displayed if the volume configuration routine is initialized at a
pressure under 1 MPa (150 psi).
3.4.4.2
CAL
PURPOSE
To calibrate the PPCK+ reference pressure transducer (RPT) and adjust the
on-board barometer. This function is considered part of PPCK+ maintenance
and is therefore covered in the maintenance section of this manual (see
Section 5.2).
PRINCIPLE
See Sections 5.2, 5.3, and 5.4.
OPERATION
See Sections 5.2, 5.3, and 5.4.
Run config?
1Yes 2No
Activate new config?
1Yes 2No
Содержание PPCK PLUS
Страница 1: ...2001 DH Instruments Inc PPCK High Pressure Controller Calibrator Operation and Maintenance Manual...
Страница 14: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 6 N NO OT TE ES S...
Страница 24: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 16 N NO OT TE ES S...
Страница 76: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 68 N NO OT TE ES S...
Страница 142: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 134 N NO OT TE ES S...
Страница 150: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 142 N NO OT TE ES S...