VENT
Purpose
Read, execute or abort a vent process.
Command
Query
“VENT
n
”
“VENT?”
Classic “VENT=
n
”
“VENT”
Arguments
n:
’1’ to start a vent process.
’0’ to abort a vent process and close the exhaust valve.
Remarks
The PPCK+ vents by generating close to atmosphere, and then opening the
exhaust valve. This program message query will return a ‘0’ if the exhaust valve is
closed, or a ‘1’ if the exhaust valve is open.
Example
(enhanced)
Cmd sent:
Query reply:
“VENT 1”
“0” (if not vented)
“1” (if vented)
Example
(classic)
Sent:
Reply:
“VENT=1”
“VENT=0” (if not vented)
“VENT=1” (if vented)
Errors
ERR# 6:
The argument is not a ‘0’ or a ‘1’.
See Also
3.1.2.2
VER
Purpose
Read the PPCK+ version.
Query “VER?”
Classic “VER”
Remarks
The software version of the PPCK+ can be read. This is useful for checking
presence of the PPCK+ and for reference purposes.
Example
(enhanced)
Query sent:
Query reply:
“VER?”
“DH INSTRUMENTS, INC PPCK+ Ver1.00 ”
Example
(classic)
Query sent:
Query reply:
“VER?”
“DH INSTRUMENTS, INC PPCK+ Ver1.00 ”
See Also
None
Содержание PPCK PLUS
Страница 1: ...2001 DH Instruments Inc PPCK High Pressure Controller Calibrator Operation and Maintenance Manual...
Страница 14: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 6 N NO OT TE ES S...
Страница 24: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 16 N NO OT TE ES S...
Страница 76: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 68 N NO OT TE ES S...
Страница 142: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 134 N NO OT TE ES S...
Страница 150: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 142 N NO OT TE ES S...