PPCK+™ Operation and Maintenance Manual
Page 75
©2001
DH Instruments, Inc
.
4.4.2
ERROR MESSAGES
The PPCK+ always replies to a program message while using the classic format. If the
program message is incorrect or contains invalid data, an error number will be returned in the
form
<ERR#n>
where n is an integer number that represents a specific error. This allows for
easy error trapping by the host computer.
In the enhanced format, each error is placed into the Error Queue as it occurs. The
<ERR?>
query can then be used to remove from the queue and reply the text description of each error
that has occurred. If you are using the COM1 port, the Error Queue is cleared each time a
program message (except
<ERR>
) is received. Section 4.5.1 explains the details of the
Error Queue. Here is a list of the possible errors numbers and the error description for each.
Table 13.
Error #’s and Descriptions
REPLY
DESCRIPTION
ERR# 0
"OK"
ERR# 2
"Text argument is too long"
ERR# 3
"User defined coefficient cannot be 0"
ERR# 4
"External device not detected"
ERR# 5
"External device improperly configured"
ERR# 6
"Numeric argument missing or out of range"
ERR# 7
"Missing or improper command argument(s)"
ERR# 8
"External device time-out error"
ERR# 9
"Unknown command"
ERR# 10
“Missing or invalid command suffix” or ”the pressure is
not greater than 1 MPa or PPCK+ is busy”
ERR# 11
"Command missing argument"
ERR# 12
"System overpressured"
ERR# 13
"Text queue overflow"
ERR# 14
"User unit not defined"
ERR# 16
“Generation failure”
ERR# 18
"Command not yet available"
ERR# 19
"Not available with absolute units"
ERR# 20
"Not available with gauge transducer”
ERR# 21
"User device not defined"
ERR# 22
"Pressure is not stable" or “Must be vented”
ERR# 23
"Option not available or installed"
ERR# 24
"Unit must be vented"
ERR# 25
"Transducer out of calibration"
ERR# 26
"COM port failed to initialize"
ERR# 27
"Internal device#1 time-out error"
ERR# 28
"Device failure"
ERR# 30
"Must be on range H3"
ERR# 31
"Upper limit is too low"
ERR# 32
"Not stable enough"
ERR# 33
"Failed to generate pressure"
ERR# 44
“Entry already exists”
ERR# 45
“Argument not allowed”
Содержание PPCK PLUS
Страница 1: ...2001 DH Instruments Inc PPCK High Pressure Controller Calibrator Operation and Maintenance Manual...
Страница 14: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 6 N NO OT TE ES S...
Страница 24: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 16 N NO OT TE ES S...
Страница 76: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 68 N NO OT TE ES S...
Страница 142: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 134 N NO OT TE ES S...
Страница 150: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 142 N NO OT TE ES S...