PPCK+™ Operation and Maintenance Manual
Page 97
©2001
DH Instruments, Inc
.
RETURN
Purpose
Start a new generation using the current target pressure.
Command “RETURN”
Classic “RETURN”
Remarks
The “RETURN” program message will start a new pressure generation using the
current settings and target pressure. Generation will continue until a new target
pressure is set, the PPCK+ goes into LOCAL mode, or an “ABORT” program
message is executed.
Example
(enhanced)
Cmd sent:
Reply:
“RETURN”
“1000.000 kPaa” (no reply if IEEE-488)
Example
(classic)
Sent:
Reply:
“RETURN”
“1000.000 kPaa”
Errors
ERR# 6:
The current target pressure is invalid.
See Also
3.2.3, 3.2.10, “PS“, “TP”
RS
Purpose Adjust
pressure at a fixed rate.
Command “RS
RateReq
”
Classic “RS=
RateReq
”
Arguments
RateReq:
The rate for the PPCK to generate to. Negative rates cause the
pressure to decrease and positive rate cause the pressure to
increase.
Remarks
The requested rate is an approximation, and the PPCK will set and maintain a rate
near to the requested rate to be the best of its ability. Actual rates that can be set
depend on the volume of the device under test. The PPCK+ will continue to maintain
the rate given until the “ABORT” command or [ESCAPE] key is used to cancel the
operation, or the UL pressure is reached.
Example
(enhanced)
Cmd sent:
Query reply:
“RS 10”
“10 kPa/s”
Example
(classic)
Sent:
Reply:
“RS=10”
“10.00 kPa/s”
See Also
“ABORT”, “RATES’, “IF”, “DF”, “IS”, “DS”, “UL”
Содержание PPCK PLUS
Страница 1: ...2001 DH Instruments Inc PPCK High Pressure Controller Calibrator Operation and Maintenance Manual...
Страница 14: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 6 N NO OT TE ES S...
Страница 24: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 16 N NO OT TE ES S...
Страница 76: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 68 N NO OT TE ES S...
Страница 142: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 134 N NO OT TE ES S...
Страница 150: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 142 N NO OT TE ES S...