PPCK+™ Operation and Maintenance Manual
©2001
DH Instruments, Inc
.
Page 86
IS
Purpose
Increase the pressure at the slow increase rate.
Command “IS
n
”
Classic “IS=
n
”
Arguments
N
‘0’
Stops the pressure increase.
‘1’
Starts the pressure increase.
Remarks
PPCK+ will increase pressure at the slow increase rate until “IS=0” or “ABORT” is
sent or the upper limit is reached.
Example
(enhanced)
Cmd sent:
Reply:
“IS 1”
“1”(no reply if IEEE-488)
Example
(classic)
Sent:
Reply:
“IS=1”
“IS=1”
Errors ERR#
6:
The
n
argument is a ‘0’ or a ‘1’.
See Also
3.1.2.2, “ABORT”, “DF”, “DS”, “IF”, “UL”, “RATES”, “RS”
LOCAL
Purpose
Returns control to the PPCK+ front panel.
Command “LOCAL”
Classic “LOCAL”
Remark
The REMOTE program message can lock the front panel out completely. The user
can return to local operation by sending the LOCAL program message, sending the
IEEE-488 ‘GTL’ command (if in enhanced format), or by cycling PPCK+ power.
Example
(enhanced)
Cmd sent:
Reply:
“LOCAL”
“LOCAL” (no reply if IEEE-488)
Example
(classic)
Sent:
Reply:
“LOCAL”
“LOCAL”
See Also
“REMOTE”
MEM
Purpose
Read the status from the power-up memory test.
Query “MEM?”
Classic “MEM”
Remarks
The PPCK+ system memory stores the user settings (units, resolution, generation
settings) and retains them when the unit is OFF. On power-up, this memory
is checked. If this memory is corrupted, all user settings are reset to default (as if the
“RESET” program message was executed), and the MEM status will be set to
reflect this.
Example
(enhanced)
Query sent:
Reply:
“MEM?”
“0” PPCK+ data corrupted and was set to factory defaults
“1” The memory was found to be OK on power-up
Example
(classic)
Sent:
Reply:
“MEM”
“MEM=0”PPCK+ data corrupted and was set to factory defaults
“MEM=1”The memory was found to be OK on power-up
See Also
3.4.4.5, “RESET”
Содержание PPCK PLUS
Страница 1: ...2001 DH Instruments Inc PPCK High Pressure Controller Calibrator Operation and Maintenance Manual...
Страница 14: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 6 N NO OT TE ES S...
Страница 24: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 16 N NO OT TE ES S...
Страница 76: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 68 N NO OT TE ES S...
Страница 142: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 134 N NO OT TE ES S...
Страница 150: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 142 N NO OT TE ES S...