PSS
Purpose
Set a new target pressure and use just the slow speed to generate the pressure.
Command “PSS
n
”
Classic “PSS=
n
”
Arguments
n:
The target pressure in the current pressure units.
Remarks
The PPCK+ will set the given target pressure using just the slow speed, and will stop
generating when the pressure has reached or passed the given target. The system
will not attempt to hold the pressure. Control stops after the pressure is set.
The “PR?”, “PRR?”, “STAT?”, or “SR?” program message can be used to monitor the
progress of the generation.
Example
(enhanced)
Cmd sent:
Reply:
“PSS 1000”
“1000.000 kPaa (no reply if IEEE-488)
Example
(classic)
Sent:
Reply:
“PSS=1000”
“1000.000 kPaa
Errors
ERR# 6:
The target pressure is out of range.
See Also
3.1.2.2, “PR”, “PRR”, “STAT”, “SR”, “PSF”, “RATES”
QPRR
Purpose
Read the last RPT pressure, rate, and ATM pressure.
Query “QPRR?”
Classic “QPRR”
Remarks
The last measured
Ready
condition, pressure rate of pressure change, and ambient
pressure for the active RPT (always Hi RPT in PPCK+) and range is returned in the
current pressure units. This program message is useful when a rapid response of
measured pressure is needed. Each data field is separated by a comma, and is
returned in the following order:
Ready/Not Ready
, pressure UNITS, rate UNITS/s,
atm UNITS.
Here are the field descriptions:
Ready/Not
Ready
:
‘R’ if the current pressure
Ready
criteria has been met,
‘NR’ if the criteria has not been met (see the “SR” program message).
pressure
The measured pressure for the selected RPT and range in the
current pressure units. This is followed by the current pressure units.
rate
The measured rate for the selected RPT and range in the current
pressure units per second. This is followed by the current
pressure units.
atm
The ambient pressure measured by the PPCK+ internal barometer in
the current pressure units (but always absolute). This is followed by
the current pressure units.
Example
(enhanced)
Query sent:
Query reply:
“QPRR?”
“R,2306.265 kPaa,0.011 kPa/s,97.000 kPaa”
Example
(classic)
Query sent:
Reply:
“QPRR”
“R,2306.265 kPaa,0.011 kPa/s,97.000 kPaa”
See Also
“PR”, “PRR”, “SR”, “ATM”, “RATE”
Содержание PPCK PLUS
Страница 1: ...2001 DH Instruments Inc PPCK High Pressure Controller Calibrator Operation and Maintenance Manual...
Страница 14: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 6 N NO OT TE ES S...
Страница 24: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 16 N NO OT TE ES S...
Страница 76: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 68 N NO OT TE ES S...
Страница 142: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 134 N NO OT TE ES S...
Страница 150: ...PPCK Operation and Maintenance Manual 2001 DH Instruments Inc Page 142 N NO OT TE ES S...