Operating Instructions Xios AE Sensor USB interface and sensors
9 Maintenance and inspection
9.3 Replace the sensor cable
67 29 433 D3780
D3780.201.01.01.02 09.2019
79
Replace the Elastomer
1.
Check the contents of the spare parts kit.
2.
Your spare parts kit contains a white elastomer frame. Carefully
place the white frame into position, flat surface facing up and
notched cutout facing the longer side of the sensor. Avoid contact
with the gold elastomer strip in the center.
After inserting the frame, apply a small amount of finger pressure
around the outer edges to ensure that the material is seated
correctly.
IMPORTANT
Use the new parts supplied with the cable for the assembly!
IMPORTANT
The spare parts for the Xios AE sensor are white. The Elastomer strip
(or frame) must be placed exactly in the slot so that the sensor works.
Attach a replacement cable
1.
Place the plug for the sensor cable properly on the sensor. Both
parts must interlock.
2.
Screw the sensor cable to the sensor using the screws supplied.
Initially only tighten the screws until a light resistance can be felt.
After this tighten both screws carefully.
3.
For each screw just tightened, remove a gel disk from its paper
backing and place it over the screw. make sure that the gel
completely covers the screw.
4.
Cover the screw heads with a new protective screw cap. Press the
cap into the plug of the sensor cable until it locks in place.
Ä
The cable replacement is now completed. The sensor can be used
again.
NOTE
Improperly tightened screws may affect sensor performance
Be sure to tighten both screws completely until they cannot be turned
any further. Inadequate tightening may create a poor connection
between the elastomer and the sensor contacts and could affect the
ability of the sensor to acquire and transfer images optimally.