![Dentsply Sirona Xios AE Скачать руководство пользователя страница 31](http://html1.mh-extra.com/html/dentsply-sirona/xios-ae/xios-ae_operating-instructions-manual_2483720031.webp)
Dentsply Sirona
Operating Instructions Xios AE Sensor USB interface and sensors
4 Operation
4.4 Slide the hygienic protective sleeve over the sensor
67 29 433 D3780
D3780.201.01.01.02 09.2019
31
4.4
Slide the hygienic protective sleeve over the
sensor
Hygienic protective sleeves for use with Xios AE sensors are available
depending on the size of the sensor (0, 1 or 2).
To reorder the hygienic protective sleeves, see "Consumables and
spare parts."
WARNING
Sensors and sensor cables must be disinfected prior to initial use.
Patients may become sick due to components that have not been
disinfected.
➢ Remove the sensor connector from the unit.
➢ Clean the sensor and the sensor cable thoroughly with
disinfectant at least twice. Refer to Hygiene.
WARNING
The hygienic protective sleeves and sensor holder tabs are single use
devices.
Patients may become sick due to unsterilized accessories.
➢ Replace the hygienic protective sleeves and sensor holder tabs
after each patient. However, they can be used multiple times on
the same patient. The adhesive on the sensor holder tabs is
suitable for gluing and detaching them from the hygienic
protective sleeve repeatedly.
➢ Under no circumstances should you slide a hygienic protective
sleeve over a sensor where a sensor holder tab is already glued
on it.
1.
Please select a hygienic protective sleeve with a size that matches
the sensor.
2.
Slide the sensor into the hygienic protective sleeve.
Ä
The hygienic protective sleeve is slightly undersized, so that it
surrounds the sensor tightly and prevents the sensor from
slipping out of place.