4 Operation
4.7 Position sensor with single-use sensor holder system
67 29 433 D3780
D3780.201.01.01.02 09.2019
47
4.7.5
Measurement exposure for endodontics
Endodontic needles and files can remain in the root canal for the
measurement exposure.
For endodontics exposures use the grey sensor holder.
C
B
A
1.
Place the grey localizer ring (B) onto the plastic guide rod (C).
2.
Place the grey sensor holder tab (A) onto the guide rod (C).
3.
Slide the sensor into the hygienic protective sleeve, see section
"Slide the hygienic protective sleeve over the sensor".
4.
Glue the sensor holder tab onto the sensor's hygienic protective
sleeve. Place the tab in the center of the sensor as shown in the
diagram.
5.
Position the sensor in the patient's mouth.
6.
Bring the X-ray tube assembly into the correct position and take an
X-ray exposure.
7.
Remove the sensor from the hygienic protective sleeve. For this
follow the instructions in section "Remove the hygienic protective
sleeve from the sensor". The used sensor holder tab and hygienic
protective sleeve must be disposed of after the examination.
8.
Clean and sterilize the guide rod and localizer ring.