Dentsply Sirona
Operating Instructions Xios AE Sensor USB interface and sensors
3 System description
3.9 Technical data
67 29 433 D3780
D3780.201.01.01.02 09.2019
25
3.9
Technical data
AE Sensor USB interface
Electric class:
Protection class II
Type of protection
against electric shock:
Applied part type BF
Degree of protection
against ingress of water:
Usual device (without protection against
ingress of water), corresponds to protec-
tion class IPX0
USB port:
Version 2.0 and 3.0
Operating voltage:
5 V DC
Device is supplied with power via the USB
connection of the PC.
Supply current:
0.5 A
Power consumption:
1.25 W
Maximum USB cable
length:
5 m
Dimensions L x W x H:
107 x 62 x 28 mm
Weight:
approx. 50 g
Xios AE Sensors
Technology:
CMOS-APS (Active Pixel Sensor)
Physical pixel size:
15 µm, image acquisition in 15 µm
Line pairs:
33.3 lp at 15 µm
Measured resolution:
28 Lp/mm
Theoretical resolution:
33.3 Lp/mm
Active sensor area:
Size 0 - sensor = 18 x 24 mm
Size 1 - sensor = 20 x 30 mm
Size 2 - sensor = 25.6 x 36 mm
External dimensions:
Size 0 - sensor = 19.4 x 28.2 x 7.5 mm
Size 1 - sensor = 21.5 x 34.2 x 7.5 mm
Size 2 - sensor = 27.1 x 40.2 x 7.5 mm
Cable length:
max. 2.70 m
Operating and transport conditions
Mode of operation:
Continuous operation
Additional information:
The unit must not be used near flammable
anesthesia gas mixtures containing air,
oxygen or nitrogen oxide.
Transport and storage
conditions:
Temperature: -25 °C (-13 °F) to 70 °C (158
°F)
Relative humidity: 20 % to 85 %
Air pressure: 500 to 1060 hPa