4 Operation
4.8 Position sensor with AimRight reusable sensor holder system
67 29 433 D3780
D3780.201.01.01.02 09.2019
55
4.8.3
Horizontal bite wing exposures
Preparing the sensor holder
For bite wing exposures use the red sensor holder.
1.
Fasten the guide rod for the sensor (C) in the perforation (B) of the
localizer ring (A).
2.
Slide the sensor into the hygienic protective sleeve, see section
"Slide the hygienic protective sleeve over the sensor".
3.
Place the sensor on the palm of your hand and clip the sensor
holder onto the sensor.
4.
Slide the sensor into the center of the sensor holder.
Position the sensor
1.
Look through the localizer ring to check the alignment of the sensor.
The sensor must be located centrally in front of the opening in the