2 Safety instructions
2.4 PC system and software
Dentsply Sirona
Operating Instructions Xios AE Sensor USB interface and sensors
12
67 29 433 D3780
D3780.201.01.01.02 09.2019
2.4
PC system and software
During the exposure, the data connection and power supply must be
ensured via the USB port. Under Power Options on the control panel
set the PC so that it never switches to stand-by operation or goes to
sleep.
Sidexis 4 may not be terminated until the end of the exposure. Before
image acquisition, close all programs that are not required for the
operation of Sidexis 4. Any programs running in the background, such
as media players, print managers, backup software, etc., may cause
Sidexis 4 to crash during the exposure. In cases of doubt, consult your
system administrator.
These operating instructions presuppose that you are familiar with the
use of the Sidexis 4 software.
2.5
Allocation of acquisition system to patient
During routine practice, the clear allocation of the acquisition system to
the patient to be examined is ensured. This also applies to the
allocation of the X-rays to the patient data stored by Sidexis 4.
2.6
Trouble-free operation
Use of this unit is permissible only if it works properly without
malfunctions. If trouble-free operation cannot be ensured, the unit must
be taken out of service, checked by authorized technicians for
malfunctions and, if necessary, repaired or replaced.
2.7
Maintenance
The owner is responsible for making sure that all inspections are
performed.
In the event that the system owner fails to fulfill the obligation to perform
scheduled inspection activities or ignores error messages, Sirona
Dental Systems GmbH and its authorized dealers cannot assume any
liability for damages.
As manufacturers of medical electrical equipment we can assume
responsibility for the safety-related features of the equipment only if
maintenance and repair are carried out only by ourselves or agencies
expressly authorized by us, and if components affecting safe operation
of the system are replaced with original spare parts upon failure.
We suggest that you request a certificate showing the nature and extent
of the work performed from those who carry out such work; it must
contain any changes in rated parameters or working ranges (if
applicable), as well as the date, the name of the company and a
signature.