Section VII User Service Information
Troubleshooting
A full color, fold-out tubing diagram of the CORE
Magellan unit is provided with this manual.
The following charts should be used when
troubleshooting CORE Magellan unit problems.
If these suggested troubleshooting procedures
do not resolve the problem, refer to the
Service
Instruction
section that follows.
The DentalEZ Technical Service Department is
available to supply service personnel with any
additional information or instructions needed to
repair or maintain the dental delivery system.
NOTICE
• Before servicing, always disconnect the
external power by unplugging the unit
from the power receptacle.
• Exercise extreme caution when
troubleshooting the electrical
components of the CORE Magellan unit.
When testing, always disconnect the
external power. When electrical power
is required, safety precautions must be
followed.
• Do not modify this equipment without
authorization from the manufacturer.
• The power module power cord is not
replaceable by service personnel.
WARNING
Delivery Head
Symptom
Possible Cause(s)
Solution
Unit will not
turn on
No air to master switch
Open stop valve in
fl
oor utility box.
Open and adjust air regulator to 80 PSI.
Check for pinched red 1/8" tubing in arms.
Faulty master switch
Turn switch on and verify air
fl
ow.
Faulty air pilot valve
Verify on air supply to top
fi
tting of valve.
If air is present at top
fi
tting of valve, verify air to foot control tubing. If
no air at tubing, valve is defective.
Unit has no
electric power
No building power
Check main circuit supply.
Faulty electric switch
Contact electrician.
Improper outlet voltage
Verify proper outlet voltage in USC.
Unit has no air
Unit is not turned on
Ensure master toggle valve is in ON position.
Air stop valve is closed
Turn stop valve on and verify 80 PSI on air gauge.
Pinched air line
Check for pinched or restricted 1/8" red tubing from
fl
oor utility box to
the master switch.
Faulty air pilot valve
Check barbs for blockage. If none, replace valve.
No water
(handpiece and
syringe)
Water toggle is in OFF position
Place the CMU switch plate toggle in ON position.
No water in clean water bottle
Fill bottle.
Pressure regulator not adjusted
Ensure clean water system gauge reads 40 PSI.
No air to bottle water system
Check for pinched or restricted airline.
Slide clamp is clamping line
Move clamp to allow water
fl
ow through line.
Air in coolant
water
Unit has not been purged
Purge water lines
(Section IV Testing, Foot Control & Handpieces)
Faulty water control valve
If unit was purged, replace water control valve.
Содержание Core
Страница 1: ...CORE TM Delivery Unit Magellan Mount User Manual ...
Страница 3: ...www DentalEZ com 866 DTE INFO ii CORETM Magellan Unit PN 2717 267C ...
Страница 11: ...www DentalEZ com 866 DTE INFO 8 CORETM Magellan Unit PN 2717 267C ...
Страница 15: ...www DentalEZ com 866 DTE INFO 12 CORETM Magellan Unit PN 2717 267C ...
Страница 31: ...www DentalEZ com 866 DTE INFO 28 CORETM Magellan Unit PN 2717 267C ...
Страница 39: ...www DentalEZ com 866 DTE INFO 36 CORETM Magellan Unit PN 2717 267C ...