www.DentalEZ.com
866-DTE-INFO
14
CORE
TM
Magellan Unit
PN: 2717-267C
Section III Installation
Magellan Unit Support
(Continued)
NuSimplicity Chair
1. Raise the base and back to full
UP
position.
2.
Disconnect
the chair from the power supply.
3. Take o
ff
the chair’s seat by removing the
1/4-20 thumb screw from underneath the seat.
Push the seat toward the back of the chair to
clear the substrate from the slide bosses, then
set the seat aside with all related hardware
(Figure 11)
.
NOTE:
If installing the CORE delivery unit on a
DentalEZ chair that is not pre-tubed, continue with
the following instructions. If installing on a pre-
tubed chair,
skip steps 4 through 10
.
Figure 11. Remove the chair seat
Substrate
1/4-20 Thumb
Screw
7. Place a level on the Magellan unit support,
parallel with and perpendicular to the chair
center line, then check the level of the unit
support.
NOTE:
If leveling is necessary, adjust the 3/16" set
screws until the unit support is level.
8. Tighten M16 bolt (55 foot-pounds).
NOTE:
If necessary, repeat steps 5, 6 and 7 after
unit and light are added to re-level.
4. Select the following items from the Magellan
unit support hardware package:
•
Three 1/2"
×
1½" Bolts
•
Three 1/2" Lock Washers
•
Four 1/2" Flat Washers
•
Four 1/2" Shims
•
Two 3/8"
×
1/2" Set Screws
5. Determine if the two outside 1/2" threaded
hole mounting pads and the Magellan unit
support are level. If the mounting pads or unit
support are not level, select the necessary
amount of 1/2" shims needed to provide a level
condition across the two pads.
6. Position the Magellan unit support so that
the three 1/2" through holes align with the
corresponding threaded holes of the chair
mount casting.
7. Install the two outside 1/2" bolts and lock
washers and the number of 1/2" shims
determined in step 5 through the Magellan
unit support and thread into the chair mount
casting.
8. Tighten both outside 1/2" bolts to allow a 3/16"
clearance at the middle 1/2" bolt location.
9. Install the middle 3/8" set screws in the
bottom of the Magellan unit support.
10. Install the middle 1/2" lock washer and 1/2"
bolt,
but
do not
fully tighten
.
Содержание Core
Страница 1: ...CORE TM Delivery Unit Magellan Mount User Manual ...
Страница 3: ...www DentalEZ com 866 DTE INFO ii CORETM Magellan Unit PN 2717 267C ...
Страница 11: ...www DentalEZ com 866 DTE INFO 8 CORETM Magellan Unit PN 2717 267C ...
Страница 15: ...www DentalEZ com 866 DTE INFO 12 CORETM Magellan Unit PN 2717 267C ...
Страница 31: ...www DentalEZ com 866 DTE INFO 28 CORETM Magellan Unit PN 2717 267C ...
Страница 39: ...www DentalEZ com 866 DTE INFO 36 CORETM Magellan Unit PN 2717 267C ...