42
COMPRESSOR USER MANUAL
866-DTE-INFO
dentalez.com
PN: 7717-004A
SECTION IV - OPERATION
TECHNICAL SCREEN (CONTINUED)
MAINTENANCE MODE > SERVICE > DEW POINT SENSOR
After replacing the dew point sensor, the current runtime will need to be reset. On the Service screen, press
DEW
POINT SENSOR
. This will open the Dew Point Sensor screen showing current runtime. Press
REPLACE
. Press
OK
to
replace dew point sensor and return to Service screen, or press
CANCEL
to cancel runtime reset and return to Dew
Point Sensor screen
(FIGURE 32)
.
Service
Current Runtime
WARNING: Replacing item will reset Current Runtime to 0 and update
the serial number (if applicable). This action CANNOT BE REVERSED.
Please confirm item name and serial number above before continuing.
20 hrs
Replace
Dew Point Sensor
Technical
Select item for service
Motor 1
Motor 2
Filter
Pack
Desiccant
Tower
Pressure
Sensor
Ambient
Sensor
Dew Point
Sensor
Service
Desiccant
Cartridge
SELECT
SERVICE
ITEM
PRESS
REPLACE
Service
Replace
Dew Point Sensor
Current Runtime
WARNING: Replacing item will reset Current Runtime to 0 and update
the serial number (if applicable). This action CANNOT BE REVERSED.
Please confirm item name and serial number above before continuing.
20 hrs
Service
Replace
Dew Point Sensor
Current Runtime
WARNING: Replacing item will reset Current Runtime to 0 and update
the serial number (if applicable). This action CANNOT BE REVERSED.
Please confirm item name and serial number above before continuing.
20 hrs
Replace Dew Point Sensor?
OK
Cancel
CANCEL
UPDATES
ACCEPT
UPDATES
FIGURE 32. DEW POINT SENSOR REPLACEMENT SCREENS
RUNTIME