3
SPECIFICATIONS
SECTION I - INTRODUCTION
AERAS 3
(C3-2)
AERAS 4
(C4-1)
AERAS 4
(C4-2)
AERAS 6
(C6-2)
AERAS 7
(C7-2)
AERAS 11
(C11-2)
Max Users
3
4
4
6
7
11
Voltage
(AC)
230
(208 - 253)
115
(103 - 126)
230
(208 - 253)
230
(208 - 253)
230
(208 - 253)
230
(208 - 253)
Power Consumption
(Amps)
10.5
23
11.5
16.3
21
32
Breaker Size
(Amps)
20
30
20
20
30
40
Fusetron Size
FRN 15
FRN 25
FRN 15
FRN 20
FRN 25
FRN 40
Number of Motor
Heads
1
2
2
2
2
3
HP Per Motor Head
2
1.25
1.25
1.25 (Left)
2 (Right)
2
2
Pressure
85 PSI/586 kPa to 115 PSI/793 kPa with ± 2% Total Error Band
High-Pressure Mode
*
105 PSI/724 kPa to 131 PSI/903 kPa
Weight
(lbs)
160
190
190
250
260
385
Dimensions
W × D × H
(in)
29 × 21 × 29
29 × 21 × 29
29 × 21 × 29
35 × 25 × 30
35 × 25 × 30
41 × 29 × 32
Dimensions
(with Sound Cover)
33.3 × 29 × 34
33.3 × 29 × 34
33.3 × 29 × 34
35 × 29 × 34
35 × 29 × 34
41.1 × 33 × 36
Peak Airflow Output
@80 PSI
(CFM)
6.0
6.6
6.6
9.3
12.0
18.1
Pump Up Cycle Time,
85 to 115 PSI
(Sec)
45
45
45
45
40
40
Tank Size
(Gal)
12
12
12
20
20
30
Filtration
(microns)
0.3 Coalescing; 0.01 Final Particulate
Dryer Duty Cycle
100%
Dryer Dew Point
(PDP)
-40°F / -40°C
dBA Levels
(with/without
Sound Cover)
65/73
70/78
70/78
72/80
76/82
78/84
NOTES:
*High-pressure mode should only be used with a compatible milling machine; not for use in general dentistry.
TRANSPORTATION AND STORAGE
Temperature range: -20°F to 165°F (-29°C to 74°C)
Relative humidity range: 0% to 95%
OPERATION
Temperature range: 35°F to 100°F (2°C to 38°C)
Relative humidity range: 0% to 95%
INDOOR USE
Altitude up to 9,000 ft (2,743 m)