REPLACE PARTICULATE FILTER
NOTE:
All compressor models use the same particulate
filter.
1.
On the control panel LCD display, go to the Technical
screen. Press and enable
MAINTENANCE MODE
(FIGURE 67)
.
NOTE:
Turning on maintenance mode disables remote
switching and puts the system in standby mode.
2. Bleed air from the system.
3. Push tab down and turn particulate filter bowl slightly
right or left to remove
(FIGURE 68A)
.
4. Remove filter by turning completely out.
5. Install new .01 micron filter
(FIGURE 68B)
.
6. Position filter bowl, push up slightly and turn left or right
into place. Tab will snap into place.
7.
On the Technical screen, press
SERVICE >
(FIGURE 69)
.
8. On the Service screen, press
FILTER PACK
(FIGURE 70)
.
MAINTENANCE (CONTINUED)
SECTION V - CARE
(Continued on the next page.)
FIGURE 67. ENABLE MAINTENANCE MODE
Service
Runtimes
Sensors
Motors
Alerts
High Pressure Mode
Maintenance Mode
07:33:14 PM
Maintenance Mode
LTE
Settings
Technical
Info
Home
Network
FIGURE 69. PRESS SERVICE >
Service
Runtimes
Sensors
Motors
Alerts
High Pressure Mode
Maintenance Mode
07:33:14 PM
Maintenance Mode
LTE
Settings
Technical
Info
Home
Network
Technical
Select item for service
Motor 1
Motor 2
Filter
Pack
Desiccant
Tower
Pressure
Sensor
Ambient
Sensor
Dew Point
Sensor
Service
Desiccant
Cartridge
FIGURE 70. PRESS FILTER PACK
FIGURE 68. REMOVE FILTER BOWL, REPLACE FILTER
B
A
FILTER
BOWL
TAB
.01
MICRON
FILTER