background image

DAC-515E-EN-BF

Defi btech DDU-100 Series

Trainer AED

User Manual

check pads

do not touch patient

analyzing

Содержание DDU-100 Series

Страница 1: ...DAC 515E EN BF Defibtech DDU 100 Series Trainer AED User Manual check pads do not touch patient analyzing...

Страница 2: ...less otherwise noted Limited Warranty The information in the Defibtech AED User Manual does not constitute any warranty as to the Defibtech AED or any related products The Limited Warranty shipped wit...

Страница 3: ...Fully Automatic 10 5 3 Available Scenarios 11 5 4 The Default Training Scenario 11 5 5 Programming the Default Scenario 12 5 5 1 Programming the Default Scenario without the Remote Control 12 5 5 2 P...

Страница 4: ...Declaration 20 10 2 Electromagnetic Emissions 20 10 3 Electromagnetic Immunity 20 10 4 Regulatory Compliance 22 11 Glossary of Symbols 24 12 Contacts 26 This User Manual provides information and oper...

Страница 5: ...ows the instructor to control training exercises and functions from a distance WARNING Do not store the Trainer AED or training accessories with a rescue AED or rescue accessories The Trainer AED and...

Страница 6: ...is in use WARNING If the Trainer AED Battery Pack and or AC Adapter becomes too hot to touch or otherwise appears faulty immediately disconnect the power cord from the AC outlet attach a Do not use o...

Страница 7: ...k for use with the Trainer AED If an unspecified battery pack or a non rechargeable battery is charged it may cause overheating explosion fire or leakage CAUTION Do not disassemble or modify the Train...

Страница 8: ...ng Battery Pack should have a life of approximately 200 charge discharge cycles The Training Battery Pack does not require a 9V battery Note If a Training Battery Pack is installed in an AED or Traine...

Страница 9: ...ining Pads are connected to a DDU 100 Series AED the AED will say Check Pads and the AED will report an error during the next automatic self test WARNING Do not store the Training Pads with a rescue A...

Страница 10: ...igure the modes without a Remote Control To switch the Trainer AED mode 1 Start with the Trainer AED off 2 Power on the Trainer AED 3 Press and release the Shift Key on the Remote Control 4 Press and...

Страница 11: ...ED will prompt the student to apply pads to the patient dummy and the instructor can then manually select rhythm simulations such as NSR and VF or any one of the above training scenarios using the Rem...

Страница 12: ...the numbers to be announced in order to advance more quickly 6 Turn the Trainer AED off The Trainer AED will now operate in the chosen training scenario until the Default Scenario is reprogrammed 5 5...

Страница 13: ...lation typically used once pads have been applied to the training dummy MOTION simulates a cardiac rhythm corrupted with excessive motion artifacts typically used once pads have been applied to the tr...

Страница 14: ...this sequence regardless of name assignment 7 1 Assigning Unique Names to Trainer AEDs To program up to four Trainer AEDs with unique names use the following procedure 1 Start with all of the Trainer...

Страница 15: ...mportant guidelines to be adhered to when cleaning the device The Training Battery Pack should be installed when cleaning the Trainer AED Do not immerse the Trainer AED in fluids or allow fluids to en...

Страница 16: ...replace Training Battery Pack Trainer AED malfunction Contact your local distributor or Defibtech representative Training Battery Pack will not recharge No power being supplied from AC source Plug AC...

Страница 17: ...of the national decree s which transpose provisions of such directive At the end of its lifetime this device can only be disposed of in compliance with the provisions of the above mentioned European...

Страница 18: ...UL 60601 1 2014 35 EU Low Voltage IEC 60601 1 2 9 2 Environmental Category Specification Operating Maintenance Temperature 0 50 C 32 122 F while operating 0 40 C 32 104 F while charging Training Batt...

Страница 19: ...mAh NiMH rechargeable battery Battery Designation 8HR 15 51 Capacity At least 5 hours of continuous operation Battery Life 2 years or 200 charge discharge cycles Typical new battery at 25 C 9 4 AC Ada...

Страница 20: ...ipment Voltage fluctuations flicker emissions IEC 61000 3 3 Not applicable Battery operated equipment 10 3 Electromagnetic Immunity Immunity test IEC 60601 test level Compliance level Electromagnetic...

Страница 21: ...ructures objects and people The ISM industrial scientific and medical bands between 150 kHz and 80 MHz are 6 765 MHz to 6 795 MHz 13 553 MHz to 13 567 MHz 26 957 MHz to 27 283 MHz and 40 66 MHz to 40...

Страница 22: ...re designed to provide reasonable protection against harmful interference in a residential installation This equipment generates uses and can radiate radio frequency energy and if not installed and us...

Страница 23: ...23 DAC 515E EN BF This page intentionally left blank...

Страница 24: ...it is OFF Turns the device OFF when it is ON Caution consult accompanying documents X X X X Consult operating instructions Refer to instruction manual booklet Do not expose to high heat or open flame...

Страница 25: ...Y Y Y Y Use by yyyy mm dd Keep dry Handle with care Transportation and storage requirements See environmental requirements on packaging LATEX Does not contain latex Product is not sterile 9 6V 1500mAh...

Страница 26: ...t Road Suite 201 Guilford CT 06437 USA Y Y Y Y Y Y Y Y Tel 1 866 333 4241 Toll free within North America 1 203 453 4507 Fax 1 203 453 6657 Email sales defibtech com Sales reporting defibtech com Medic...