5.3 Available Scenarios
Six training scenarios (described below) are available for training exercises and
can be changed during training as desired.
The following scenarios are available:
1. VF (Ventricular Fibrillation) that converts to a non-shockable rhythm
(normal sinus rhythm) after the 1st shock. This scenario is based
on the American Heart Association AED training scenario and
allows for a quick overview and demonstration of AED operation.
2. Non-shockable rhythm throughout.
3. Bad pads indication until the pads are disconnected and then
re-connected (simulating replacement of the pads), followed by
VF that converts after the 1st shock to a non-shockable rhythm
(normal sinus rhythm).
4. VF that converts on the 2nd shock to a non-shockable rhythm
(normal sinus rhythm).
5. Non-converting persistent VF.
6. Pads not applied to the patient. This scenario is recommended as
the Default Scenario when using the Remote Control. The Trainer
AED will prompt the student to apply pads to the patient dummy,
and the instructor can then manually select rhythm simulations
(such as NSR and VF) or any one of the above training scenarios
using the Remote Control.
Note:
When the Trainer AED is turned on, it will power up in the Default
Scenario, not in the last-used scenario (see Section 5.5 for instructions on
setting the Default Scenario).
Note:
Training scenarios 1-5 assume that pads have already been applied to
the patient if pads are connected to the Trainer AED when the Trainer AED is
turned on. For properly sequenced exercises using these training scenarios, the
student should either apply the pads to the patient before turning the unit on, or
turn the unit on with pads disconnected, apply the pads to the patient, and then
connect the pads to the Trainer AED.
5.4 The Default Training Scenario
When the Trainer AED is first turned on, it operates in the “Default Training
Scenario.” The Default Training Scenario can be changed to any of the six
available training scenarios.
Note:
New Trainer AEDs are factory programmed to initially start with the
Default Training Scenario set to scenario #6.