10 Electromagnetic Conformity
10.1 Guidance and Manufacturer’s Declaration
DDU-100 Series Trainer AEDs are intended for use within the electromagnetic
environment specified below. The customer or the user of the DDU-100 Series
Trainer AED should assure that it is used within the stated environmental
specifications.
10.2 Electromagnetic Emissions
Emissions test
Compliance
Electromagnetic environment –
guidance
RF emissions
CISPR 11
Group 1, Class B
The DDU-100 Series Trainer
AED uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
Harmonic emissions IEC 61000-3-2
Not applicable
Battery operated equipment
Voltage fluctuations/flicker emissions
IEC 61000-3-3
Not applicable
Battery operated equipment
10.3 Electromagnetic Immunity
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment – guidance
Electrostatic discharge (ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±6 kV contact
±8 kV air
There are no special
requirements with
respect to electrostatic
discharge.
Electrical fast transient/burst
IEC 61000-4-4
Not applicable
Not applicable
Battery operated
equipment
Surge
IEC 61000-4-5
Not applicable
Not applicable
Battery operated
equipment
Voltage dips, short
interruptions and voltage
variations on power supply
input lines IEC 61000-4-11
Not applicable
Not applicable
Battery operated
equipment
Power frequency (50/60 Hz)
magnetic field IEC 61000-4-8
3 A/m
3 A/m
Power frequency
magnetic fields should
not be greater than
levels characteristic of
a typical location in a
commercial or hospital
environment.