6
DAC-515E-EN-BF
2.1 Warnings
WARNING
The Trainer AED and training accessories cannot be used to
defibrillate patients.
WARNING
All training instructors must read these instructions for use
before using the Trainer AED.
WARNING
Do not use in the presence of flammable gases or
anesthetics.
WARNING
Only use the provided AC Adapter to charge the Training
Battery Pack. Using other AC adapters may cause unsafe
conditions.
WARNING
Do not charge a battery pack which is damaged, discolored
or has leakage. It may cause overheating, explosion or fire.
WARNING
Do not cover up the Trainer AED when it is in use.
WARNING
If the Trainer AED, Battery Pack, and/or AC Adapter becomes
too hot to touch or otherwise appears faulty, immediately
disconnect the power cord from the AC outlet, attach a “Do
not use” or “To be repaired” label to the item(s) and contact
your local distributor or Defibtech representative.
WARNING
Do not overcharge the Trainer AED Battery Pack.
WARNING
Do not store the Trainer AED or training accessories with
a rescue AED or rescue accessories. The Trainer AED and
training accessories cannot provide therapy.
WARNING
If the Training Battery Pack is mistakenly inserted in a DDU-
100 series AED, the AED cannot be used to perform a
rescue. Replacing the Training Battery Pack with a rescue
battery pack will immediately allow the DDU-100 Series
AED to be used to perform a rescue.