10
DAC-515E-EN-BF
5 Training Scenarios
5.1 Trainer AED Modes
(Semi-Automatic and Fully Automatic)
New Trainer AEDs are factory programmed to function in Semi-Automatic
mode, but can be configured to operate in either Semi-Automatic or Fully
Automatic mode. All of the training scenarios noted in Section 5.3 can be used
in either mode.
Note:
Please ensure that the Trainer AED is configured to the Trainer AED mode
for which training is intended.
5.2 Configuring the Trainer AED Mode
(Semi-Automatic or Fully Automatic)
Note:
To switch between the Semi-Automatic and Fully Automatic Trainer AED
modes you must have a Remote Control. It is not possible to configure the
modes without a Remote Control.
To switch the Trainer AED mode:
1. Start with the Trainer AED off.
2. Power on the Trainer AED.
3. Press and release the
Shift
Key on the Remote Control.
4. Press and release the
Alt
Key on the Remote Control.
5. Press and release the
Pause
Key on the Remote Control.
6. The Trainer AED will announce “Training Mode” for Semi-Automatic
mode or “Version 2 Training Mode” for Fully Automatic mode.
The Trainer AED mode has now been programmed and the Trainer AED will
start using this mode every time the unit is turned on. Each time the unit is
turned on, the unit will announce to which training mode it is programmed.
Note:
The Trainer AED mode is stored in the Training Battery Pack, therefore if
the Training Battery Pack is moved to another unit, the Trainer AED mode will be
that of the Training Battery Pack used in the unit.