3 Training Components
3.1 Trainer AED Unit
WARNING
The Trainer AED and training accessories cannot be used to
defibrillate patients.
The components required to operate the Trainer AED are:
1. Trainer AED (DDU-100TR series).
2. Training Battery Pack (DBP-RCx series).
3. Training Pads - 1 adult set (DDP-101TR) supplied.
4. Remote Control (DTR-4xx series) - supplied, but use is optional.
3.2 DBP-RCx Series Training Battery Pack
Before use, the Training Battery Pack should be charged using the supplied
battery pack charger (DTR-2xx series). The battery will fully charge in 12-14
hours.
Note:
Avoid overcharging for longest battery life. When proper charging
procedures are followed, the Training Battery Pack should have a life of
approximately 200 charge/discharge cycles.
The Training Battery Pack does not require a 9V battery.
Note:
If a Training Battery Pack is installed in an AED or Trainer AED with a 9V
battery, the ASI indicator light in the upper right corner of the AED will blink
red and a warning chirp will be generated to warn the user that the AED is not
capable of operating as a rescue defibrillator.
WARNING
Only use the provided AC Adapter to charge the Training
Battery Pack. Using other AC adapters may cause unsafe
conditions.
WARNING
If the Training Battery Pack is mistakenly inserted in a DDU-
100 series AED, the AED
cannot
be used to perform a
rescue. Replacing the Training Battery Pack with a rescue
battery pack will immediately allow the DDU-100 Series AED
to be used to perform a rescue.
WARNING
Do not store the Training Battery Pack with a rescue AED.