page 28
IVPC Software Reference
DS-100C Operating Manual
and voltage values are known to be true zeros. This command is typically used
for calibration purposes.
Format
“Z,n1,n2”
n1 = Voltage gain code 0,1,2, or 3.
n2 = Current gain code 0,1,2, or 3.
Gain Codes
0 = 1X
used for 600 V, not used for current
1 = 10X
used for 60 V, 100 A
2 = 100X
used for 6 V, 10 A
3 = 1000X
not used for any measurements
Example
command “Z,0,1”
response “*”
response “>”
Pre-Curve “E”
This command initiates the pre-curve sequence. The sequence has up to a 5
second discharge of the tracer load capacitors followed by a 7 second pre-
charge.
Format
“E”
Start pre-curve sequence.
Example
command “E”
response “*”
7 to 12 second delay
response “>”
Содержание DS-100c
Страница 8: ......