page 38
Trouble Shooting
DS-10/125 Operating Manual
7.0 Troubleshooting
•
Data is not plotted after taking an I-V curve.
-
Proper connection has not been made to PV system (check
DISCONNECT switch).
•
I-V curve is flat.
-
PV short circuit current exceeds current range (check current range
setting). Check for bad connections to PV module or array.
•
I-V curve does not start at 0 V (does not reach I
sc
).
-
Pre-charge fuse has blown (check and replace).
•
DS-Tracer Stops Working (Locks Up)
-
Battery out of power (charge battery).
-
Unknown Reasons (reset DS-Tracer by recycling power).
•
Cannot establish connection of external computer to DS-Tracer.
-
DS-Tracer is not turned on or has turned itself off.
-
DS-Tracer RS-232 or USB port is not connected to computer.
-
IVPC is set to wrong COM port.
•
DS-Tracer will not accept external command.
-
Parameters are not properly formatted.
•
Irradiance readings are incorrect.
-
Pyranometer coefficients are incorrect.
-
Analog circuit requires factory calibration