page C14 

 

DS-100C Operating Manual 

 

connecting or disconnecting the test leads, it becomes more important at higher 

PV voltages. 

 

Note: the DS-Tracer DISCONNECT switch is not rated to break the electrical 
current at the full voltage and current rating of the DS-Tracer. 

 

The DS-Tracer should always be connected through a PV system disconnect 

rated at the PV system V

oc

 and I

sc

 values as shown in Figure 9. A system switch 

provides a way to break a circuit should the DS-Tracer have a major internal 

failure. 

 

Figure 9. Use of PV system disconnect switch. 

Voltage Range Switch 

The voltage-range switch changes the electrical configuration of the capacitors 

that form the load for the PV array. It will not affect the measurement range of 

the I-V curve but it does affect how quickly an I-V curve is taken. In the low 

voltage setting, the DS-Tracer has more load capacitance and therefore a curve 

is swept slower. Note: This affects how quickly the PV array sweeps from I

sc

 to 

V

oc

. It does not affect how long it takes a curve to plot in IVPC.  

The higher I

sc

 and the lower V

oc

 the more quickly a curve is taken. If curves are 

swept too fast, the dynamic characteristics of the PV array can become 

DS-Tracer 

 

 

PV Array 

System 

Disconnect 

Switch 

 

Test Lead 

Содержание DS-100c

Страница 1: ...ACER User Manual Daystar Inc 3240 Majestic Ridge Version 6 02 Las Cruces NM 88011 November 2011 575 522 4943 pvvern zianet com http www daystarpv com daystar inc daystar inc daystar inc E n e r g y E...

Страница 2: ...This I V curve tracer was built to Daystar Inc specifications by Raydec Inc 8210 La Mirada Suite 700 Albuquerque New Mexico 87109...

Страница 3: ...3 Starting IVPC 4 Taking an I V Curve 5 If You Get an Error 9 Displaying Saved Curves 10 Summary of Steps for Taking IV Curves 11 2 0 Panel Layout 12 Pre Charge Fuse 13 PV Voltage Sense Connectors 13...

Страница 4: ...Irradiance Connections 25 Multiple Current Ranges 26 External Command Set 26 Read I V B 27 Auto Zero I V Z 27 Pre Curve E 28 Read Analog Input R 29 PV V I Scales Factors S 31 Take I V Curve T 33 Read...

Страница 5: ...ting Manual page iii PV Voltage Measurements 44 PV Current Measurements 44 RANGE switch settings 44 Temperature Measurement 44 Irradiance Measurement 44 Power Supply 45 Load Capacitors 45 Approximate...

Страница 6: ...NG PRECAUTIONS ARE NOT FOLLOWED THIS INSTRUMENT MUST ONLY BE USED FOR TAKING ELECTRICAL CURRENT AND VOLTAGE I V CURVES OF PHOTOVOLTAIC PV SOLAR MODULES THIS INSTRUMENT MUST NOT BE USED BEYOND ITS OPER...

Страница 7: ...aximum current a photovoltaic cell can generate with the output terminals shorted I V Current Voltage PV Photovoltaic V Voltage VHIGH High voltage range 600 Volts VLOW Low voltage range 150 Volts Voc...

Страница 8: ......

Страница 9: ...erence Manual The computer running the IVPC application may be connected to the DS Tracer by RS 232 serial or USB Most DS Tracer operations are performed with the IVPC program Refer to the IVPC manual...

Страница 10: ...s The DS Tracer battery should be fully charged The DS 100C has a universal power supply that works with AC line voltages from 85 Vac to 264 Vac 50 60 Hz The DS Tracer DISCONNECT switch should be in t...

Страница 11: ...9 pin connector as shown in Figure 8 The standard USB connection is labeled Main USB Either of these connections may be used depending on the configuration of the computer running the IVPC program Th...

Страница 12: ...h opened Starting IVPC The following description assumes the use of IVPC 2 X IVPC opens to the Data Grid window shown in Figure 1 IVPC Data Grid window The Data Grid is a spread sheet that lists all o...

Страница 13: ...ld be connected to the DS Tracer and running IVPC From the File menu shown in Figure 2 select New I V Curve or press Ctrl N control and N keys at the same time Figure 2 Select New Curve in the File me...

Страница 14: ...up to 12 seconds to initialize the tracer hardware Typically you will enter curve data during this delay For example if not using the Auto feature to read an irradiance or temperature channel you cou...

Страница 15: ...he status message changes and the Take Curve button is enabled as shown in Figure 4 The status message indicates that a curve can be taken by using the mouse to press the Take Curve button or by press...

Страница 16: ...Curve the DS Tracer takes approximately 6 seconds to take the curve and plot the data as shown in Figure 5 You may save this data by selecting Save I V Curve from the File menu The curve is stored in...

Страница 17: ...will power down and you will receive an error indicating that the DS Tracer does not respond Simply turn the DS Tracer switch off wait several seconds and turn it back on Allow the tracer approximatel...

Страница 18: ...curves are selected from the Data Grid The Data Grid displays the curve files in the current directory Be sure to set the current working directory to the sub directory where the curves are stored Use...

Страница 19: ...to connect the tracer to the PV system through a system disconnect switch rated for Isc and Voc Refer to the discussion of the DISCONNECT switch in the section 2 0 Panel Layout 6 Set the voltage rang...

Страница 20: ...SCONNECT VOLTAGE SENSE daystar inc Manufacturedby Raydec Inc Albuquerque NM E N E R G Y E N G I N E E R I N G daystar daystar FULL CHARGE TURN DISCONNECTSWITCH OFF BEFORE CONNECTING OR DISCONNECTING P...

Страница 21: ...sconnected from the PV array PV Voltage Sense Connectors When an I V curve is taken the load current flows through the test lead into the large red and black PV connectors The red and black connectors...

Страница 22: ...PV system disconnect switch Voltage Range Switch The voltage range switch changes the electrical configuration of the capacitors that form the load for the PV array It will not affect the measurement...

Страница 23: ...DS Tracer power This switch should be on before connections are made to a PV array The curve tracer takes approximately 10 seconds to initialize after it is turned on Be sure to allow for this before...

Страница 24: ...acer functions are controlled with the IVPC software but you can create your own software using the DS Tracer command codes described in section 6 0 DS 100C Additional Capabilities The DS Tracer uses...

Страница 25: ...Appendix C DS Tracer Technical Specifications for details on the voltage range of the irradiance inputs These inputs are differential but there is a 10 K resistor form the negative black input and the...

Страница 26: ...d transfers the data set to the control computer The data set includes the I V curve data plus values of open circuit voltage Voc and short circuit current Isc Peak power voltage at peak power Vpk cur...

Страница 27: ...m and the load capacitors SCRs are susceptible to false triggering turn on if the voltage across them is increased too rapidly The DS Tracer DISCONNECT switch uses two 1M resistors one each for the po...

Страница 28: ...it is best to normalize this rate to an individual PV cell This can be done by dividing the rate by the number of PV cells in series If the previous example had fifty cells in series the rate of each...

Страница 29: ...power or a DC source such as a PV module The DS 100C has a universal AC power supply and therefore will work with single phase AC line voltages from 85 Vac to 264 Vac 50 60Hz To charge the battery fr...

Страница 30: ...utomatic charging system maintains the battery at its peak charge level and will not overcharge the battery When the battery is fully charged the full charge indicator comes on You do not need to turn...

Страница 31: ...surges or internal faults can cause this fuse to fail Return the DS Tracer for repairs if this fuse blows repeatedly Replace this fuse with the type indicated on the DS Tracer panel A user accessible...

Страница 32: ...systems with many modules in series You can avoid damage to a thin film module by inserting a by pass diode across the module terminals as shown in Figure 10 This diode limits the reverse voltage to a...

Страница 33: ...inputs when taking I V curves Temperature Connections Two standard thermocouple connectors are available on the front of the DS 100C A standard DS 100C comes with Type T thermocouple connectors In ord...

Страница 34: ...S 232 serial or USB control ports The DS 100C sends a prompt character when it is ready to receive a command line Commands should only be sent to the DS Tracer after receiving this prompt Each command...

Страница 35: ...arameter is passed see below Format B n1 n2 Z n1 Voltage gain code 0 1 2 or 3 n2 Current gain code 0 1 2 or 3 Z Use current offset values see auto zero Gain Codes 0 1X used for 600 V not used for curr...

Страница 36: ...Gain Codes 0 1X used for 600 V not used for current 1 10X used for 60 V 100 A 2 100X used for 6 V 10 A 3 1000X not used for any measurements Example command Z 0 1 response response Pre Curve E This co...

Страница 37: ...o the circuits used for the current and voltage readings The analog channels refer to the irradiance pyranometer and temperature readings as well as auxiliary analog port and internal readings such as...

Страница 38: ...l reference The option 2 format does not return any values Option 3 sets the averaging period for all analog readings The default value is 16 667 microseconds This is equal to 1 60 Hz ac power cycle A...

Страница 39: ...onstant for each gain code The integer values returned by the DS Tracer for the PV voltage and current values are multiplied by these constants to convert these values to engineering units The constan...

Страница 40: ...page 32 IVPC Software Reference DS 100C Operating Manual 4 byte 32 bit voltage constant 4 byte 32 bit current constant response...

Страница 41: ...mmand X Format T L Use low current range T H Use high current range Example 1 command T L response response Take low current curve Example 2 command T H response response Take a high current range cur...

Страница 42: ...e 9600 RS 232 port baud rate response 9600 secondary port baud rate response 600 nominal high voltage range response 150 nominal low voltage range response 100 nominal high current range response 10 n...

Страница 43: ...t NA response 10 0839 10A range current constant response 100 0529 100A range current constant response 1000 0 1000A current constant NA response 4 00183E 3 5 mV analog constant response 4 00381E 2 50...

Страница 44: ...turns itself off after 15 minutes without any activity This command enables this feature By default this feature is enable This command is only necessary if the power down feature has been disabled u...

Страница 45: ...ge data int Amps 256 Array of current data float Vscale Voltage constant float Iscale Current constant float Temp1 Temperature 1 float Temp2 Temperature 2 float Irrad1 Irradiance 1 float Irrad2 Irradi...

Страница 46: ...reach Isc Pre charge fuse has blown check and replace DS Tracer Stops Working Locks Up Battery out of power charge battery Unknown Reasons reset DS Tracer by recycling power Cannot establish connectio...

Страница 47: ...ion is added to this file format F curve name D date T time S site B sub system M module P temperature 1 Q temperature 2 R irradiance 1 U irradiance 2 X miscellaneous H short circuit current O open ci...

Страница 48: ...255 ERROR 15 COMMAND LINE OVERFLOW A command has been sent to the DS Tracer before the previous command has finished Wait for return prompt ERROR 16 PARAMETER TOO LONG The maximum length of any comma...

Страница 49: ...onnection ERROR 40 DISCONNECT SWITCH IS OFF The DS Tracer is trying to take an I V curve but the DS Tracer DISCONNECT switch is off ERROR 50 INVALID NUMERIC PARAMETER Parameter must represent valid nu...

Страница 50: ...ort number outside this range even if a COM port number from 1 to 4 is not in use Use the Windows OS device manager to change the COM port number assigned to this virtual COM port Often Windows will w...

Страница 51: ...are has been connected and begin the process of installing the driver The instructions on the CD ROM provide much more detail but in general you will want to have Windows use the drivers in the driver...

Страница 52: ...c Accuracy Larger of 0 5 of reading or 2 8 mV Resolution 0 18 mV PV Current Measurements Gain Code 1 Range 100 A dc Accuracy Larger of 0 5 of reading or 45 m A Resolution 3 mA Gain Code 2 Range 10 A d...

Страница 53: ...Hawker Energy Products 0809 0020 Hours of Use Up to 13 Hours on a full charge Charge Time Less than 8 hrs Load Capacitors Low Voltage Range Capacitance 5 000 uf Voltage Limit 350 V dc High Voltage Ran...

Отзывы: