Advanced Entry Calibration (continued)
Step
Action
Displayed Message
Comment
17
After all entries have been
made, press the
Key
to back out of the menu and
then press the
Key
to save the settings and
return to the Ready Mode.
SAVE CHANGES?
ENTER KEY = YES
From READY, press the
FEED Key to advance to
the next label TOF.
Note: If the Advanced Entry Calibration fails, enter MEDIA SETTINGS / CALIBRATION /
ADVANCED ENTRY / SENSOR GAIN and lower the selected Gain Number by one.
Select the new Gain Number then save the changes and exit the menu. Retest
your media. If unsuccessful, repeat the procedure until a usable gain is
obtained.
5.3 Reset Methods
There are three reset methods, each with a different affectivity. Proceed according to the
desired result:
5.3.1 Soft Reset
To clear any temporary host settings, press and hold the CANCEL Key for approximately
four seconds.
5.3.2 Level One Reset
To return the factory default settings or (if saved) a Factory Setting File, turn OFF the
printer. Press and hold the PAUSE and CANCEL Keys while turning ON the printer and until
the SYSTEM RESET message flashes.
Note: The same as selecting SET FACTORY DEFAULTS; see Section 4.2.5.
5.3.3 Level Two Reset
To return the firmware default settings and clear all parameters, turn OFF the printer. Press
and hold the PAUSE, FEED, and CANCEL Keys while turning ON the printer and until the
SYSTEM RESET message flashes.
Note: Calibration is required; see Section 5.2.