TECHNICAL SUPPORT
82
VLASE™ IR
TROUBLESHOOTING
If a problem occurs during operation, first check the following troubleshooting. If you
cannot fix the problem, contact Datalogic customer service (see “Technical Support” on
List of warning and error states
POSSIBLE
CAUSE
ACTION
OUTPUT STATE
COMMAND BOX
CONNECTOR
Warning invalid start
sequence
Incorrect turning-on
sequence
Set Key and Enable Switchs to
“0” and repeat the turning-on
sequence. See “Turning on
sequence” on page 59
SYSTEM_ALARM
POWER_ON
ENABLE_OUT
OFF
OFF
OFF
System Error
Interlock error
- Check Interlock connector sig-
nals (see “Interlock” on page 23)
- Check Interlock connector
presence
SYSTEM_ALARM
POWER_ON
ENABLE_OUT
ON
OFF
OFF
- Check if Command Box con-
nector is present
- Check CONNECTOR_PRES-
ENCE input signal of the Com-
mand Box connector (see
“Command Box (laser control)”
on page 27)
Temperature error
Check the temperature of the
environment where the control
rack is placed. The temperature
must not exceed the operative
limit
Master Oscillator
error
Contact Datalogic Technical
Support
SYSTEM_ALARM
POWER_ON
ENABLE_OUT
ON
OFF
OFF
STATUS LED
DESCRIPTION
Blinking ORANGE
Blinking RED
Содержание VLASE IR 1109-1 42 Series
Страница 1: ...VLASE IR USER MANUAL InfraRed Laser Marker ...
Страница 72: ...USE AND OPERATION 64 VLASE IR ...
Страница 98: ...LABELS 90 VLASE IR Positioning of labels on the resonator Figure 2 External labels rsonator location ...
Страница 101: ...PERFORMANCE LEVEL PL USER MANUAL 93 BLOCK DIAGRAM ...
Страница 121: ...THERMALIZATION AND SUPPRESSION OF GIANT PULSES USER MANUAL 113 ...
Страница 133: ...USER MANUAL 125 APPENDIX G MECHANICAL DRAWINGS ...
Страница 134: ...MECHANICAL DRAWINGS 126 VLASE IR CONTROL RACK NOTE Please refer to Datalogic website for detailed drawings ...
Страница 135: ...RESONATOR USER MANUAL 127 RESONATOR NOTE Please refer to Datalogic website for detailed drawings ...