TROUBLESHOOTING
USER MANUAL
83
List of problems related to laser marker states
POSSIBLE CAUSE
ACTION
Blinking
GREEN
- Lighter™ Suite marking SW
corrupted
- C:\ or D:\ drive corrupted
Restore the laser marker using USB recov-
ery disk (see “How to recover the laser
marker” on page 120)
Steady
GREEN
Key Switch selector in wrong
position
Check the Key Switch is set to 1 (see “Turn-
ing on sequence” on page 59)
EXT_KEY contact is LOW level
or disconnected
Check EXT_KEY input signal on the Com-
mand Box connector is set to HIGH level
(see “Command Box (laser control)” on
page 27)
Steady
ORANGE
Enable Switch selector in
wrong position
Check the Enble Switch is set to 1 (see
“Turning on sequence” on page 59)
EXT_ENABLE_A and/or
EXT_ENABLE_B contact are
LOW level or disconnected
Check EXT_ENABLE_A and EXT_EN-
ABLE_B input signals on the Command Box
connector are set to HIGH level (see “Com-
mand Box (Laser Control)” on page 35)
List of most common problems
PROBLEM
DESCRIPTION
STATUS LED
Laser marker never goes to
WAIT FOR START state
Laser marker never goes to
STANDBY SHUTTER CLOSED
state
Laser marker never goes to
READY state
PROBLEM DESCRIPTION
POSSIBLE CAUSE
ACTION
Laser marker doesn't turn ON
Power supply cable disconnected
Check the POWER SUPPLY CABLE con-
nection to the power grid
Power Suppply SWITCH set to OFF
Check that the POWER SUPPLY switch is
set to ON
FUSES damaged
Check that FUSES of power supply are not
damaged
Impossible to control the marking process
using the Command Box control signals
Incorrect integration
Check that the Command Box control sig-
nals are compatible with the external
devices used for integration
Bad marking result
F-Theta scan lens is dirty
Clean the F-Theta scan lens. See “F-Theta
scan lens cleaning procedure” on page 80
Incorrect focus distance
Check that the material to be marked is
placed at the right working distance. See
“Marking Area Specification” on page 28
Incorrect laser parameters
Check that the laser parameters set in the
layout are appropriate for the material to
be marked. Contact Datalogic Technical
Support
Laser marker doesn't engrave
Incorrect laser marker state
Check the laser marker state is set to
READY
F-Theta lens protection not removed
Check that the F-Theta scan lens protec-
tion has been removed
The optical path is obstructed
Check that the optical path is free
Содержание VLASE IR 1109-1 42 Series
Страница 1: ...VLASE IR USER MANUAL InfraRed Laser Marker ...
Страница 72: ...USE AND OPERATION 64 VLASE IR ...
Страница 98: ...LABELS 90 VLASE IR Positioning of labels on the resonator Figure 2 External labels rsonator location ...
Страница 101: ...PERFORMANCE LEVEL PL USER MANUAL 93 BLOCK DIAGRAM ...
Страница 121: ...THERMALIZATION AND SUPPRESSION OF GIANT PULSES USER MANUAL 113 ...
Страница 133: ...USER MANUAL 125 APPENDIX G MECHANICAL DRAWINGS ...
Страница 134: ...MECHANICAL DRAWINGS 126 VLASE IR CONTROL RACK NOTE Please refer to Datalogic website for detailed drawings ...
Страница 135: ...RESONATOR USER MANUAL 127 RESONATOR NOTE Please refer to Datalogic website for detailed drawings ...