![DAKOTA ULTRASONICS MiniMax v2.0 Скачать руководство пользователя страница 82](http://html1.mh-extra.com/html/dakota-ultrasonics/minimax-v2-0/minimax-v2-0_operation-manual_3268519082.webp)
Dakota Ultrasonics
76
Note:
The two-point option allows the user to use two separate points for the
probe zero calibration.
Note:
The one point option can optionally be selected if only one cal bar will be
used. Although the procedures below are described with the two-point option in
mind, they will also work in the case of a one-point calibration.
4) Press the
and
arrow keys multiple times to scroll through the sub
menu items until
MEASURE ZERO
is highlighted.
5) Press the
key to display the
Two Point Zero
list box.
6) Press the
and
arrow keys multiple times to scroll through the
Two
Point Zero
list until
PHY LEN 1
is highlighted.
Note:
PHY LEN 1 is the actual physical length of the first calibration standard
noted on the certificate included with the bars. If the glass block is being used,
the 1” side should be used for PHY LEN 1.