![DAKOTA ULTRASONICS MiniMax v2.0 Скачать руководство пользователя страница 68](http://html1.mh-extra.com/html/dakota-ultrasonics/minimax-v2-0/minimax-v2-0_operation-manual_3268519068.webp)
Dakota Ultrasonics
62
obtaining the information needed to use the standard bolt for a probe zero calibration
bar.
Setting up a Bolt Standard – For the First Time
1) With the menu items already active, press the
key multiple times to tab
right and the
key multiple times to tab left until the
CAL
menu is
highlighted and displaying the submenu items.
2) Press the
and
arrow keys multiple times to scroll through the sub
menu items until
ZERO MODE
is highlighted.
3) Press the
and
arrow keys multiple times to scroll the zero mode
options until the
FIXED
option is displayed.
Note:
Each time the arrow keys are pressed a confirmation screen will appear
to warn the user that the zero mode is being changed. Press the
key to
advance to the next mode option.
Note:
The fixed option sets the delay factor to zero. In other words, disables
the zero feature altogether.
4) Press the
key multiple times to tab right and the
key multiple times
to tab left until the
MATL
menu is highlighted and displaying the submenu
items.
Содержание MiniMax v2.0
Страница 2: ......
Страница 49: ...MiniMax Bolt Tension Monitor 43 Figure 6 Reflection in a bending bolt...
Страница 192: ...Dakota Ultrasonics 186 4 Once the desired language is displayed press the key to return to the measure screen...