![DAKOTA ULTRASONICS MiniMax v2.0 Скачать руководство пользователя страница 132](http://html1.mh-extra.com/html/dakota-ultrasonics/minimax-v2-0/minimax-v2-0_operation-manual_3268519132.webp)
Dakota Ultrasonics
126
RF
The RF mode shows the waveform in a similar fashion to an oscilloscope. It shows
both the positive and the negative peaks. The peak (either positive or negative)
selected for measurement is shown in upper portion of the display.
Rectified
The rectified A-Scan mode shows half of the waveform. Either the positive or
negative peaks are displayed, depending on the polarity selected.
Digits
The Digits view is a basic digital look and feel. The larger digits make it much easier
for the operator to monitor the measurements. The Limits Bar has also been added
to the Digits view to provide the user with yet another visual tool for easily monitoring
if the fastener is inside specific tolerances.
Note:
For both RF and Rectified views, it’s important to note that a measurement
must fall inside the displays viewable range in order to see the waveform. However,
even if the waveform is outside of the viewable range of the display, a measurement
can be taken and viewed on the digital readout. If the wave is outside of the display
range, you can manually change the range by adjusting the Delay and Width values,
or you can use the Auto Find feature located in the UTIL menu of the tabbed menu
items.
The following is a reference list of the viewable features as shown above:
A.
Repeatability/Stability Indicator
– Indicates the reliability of the measured
bolt length. When all the vertical bars are fully illuminated and the last digit on
the digital thickness value is stable, the
MiniMax
is reliably measuring the
same value on a consistent basis.
B.
Battery Icon
– Indicates the amount of battery life the
MiniMax
has
remaining.