![DAKOTA ULTRASONICS MiniMax v2.0 Скачать руководство пользователя страница 128](http://html1.mh-extra.com/html/dakota-ultrasonics/minimax-v2-0/minimax-v2-0_operation-manual_3268519128.webp)
Dakota Ultrasonics
122
2) Press the
and
arrow keys multiple times to scroll through the sub
menu items until
LOAD CAL CALC
is highlighted.
3) Press the
key to run the calculation and display the confirmation
screen.
Note:
The regression or vector coefficient will be displayed. The closer this
number is to 1, the better the fit, or least amount of error.
4) Press the
key to accept the calculation, or the
key to cancel and
return to the
CAL
menu.
5) Press the
key to return to the measurement screen.
Note:
The new Load Factor and Offset values can be reviewed and edited in
the GEOM tabbed menu item.
6) Calibration is completed.