![DAKOTA ULTRASONICS MiniMax v2.0 Скачать руководство пользователя страница 145](http://html1.mh-extra.com/html/dakota-ultrasonics/minimax-v2-0/minimax-v2-0_operation-manual_3268519145.webp)
MiniMax
Bolt Tension Monitor
139
The procedures to adjust the
Thresholds
are outlined below:
Adjusting the Thresholds using the Hot Menus
1) Press the
key once to activate measure menu items. Press the
key multiple times to move right and the
key multiple times to move left,
until the desired
Threshold
cell is highlighted.
2) Press the
and
arrow keys multiple times to scroll the highlighted
value.
3)
Alternatively
, press the
key to display the
Digits Edit Box
.
4) Press the
and
arrow keys multiple times to scroll the highlighted
value.
5) Press the
and
arrow keys multiple times to scroll the digit
locations.
6) Repeat steps
4 & 5
until the
Threshold
value is correctly displayed.
7) Press the
key to return to the measurement screen, or
to cancel
entering the
Threshold
.
The user can also access and adjust the
Threshold
from the tabbed menus.
However, this method is more tedious than making the adjustments using
the Hot
Menus
. The procedure using the tabbed menus is outlined below:
Adjusting the Threshold using the Tabbed Menus