![DAKOTA ULTRASONICS MiniMax v2.0 Скачать руководство пользователя страница 60](http://html1.mh-extra.com/html/dakota-ultrasonics/minimax-v2-0/minimax-v2-0_operation-manual_3268519060.webp)
Dakota Ultrasonics
54
1) Press the
and
arrow keys multiple times to scroll through the new
Group List Items
until
CREATE NEW GROUP?
is highlighted.
2) Press the
key to accept the group parameters, and activate the
confirmation screen.
3) Press the
key to save the
New Group
, or the
key to cancel
the
New Group
setup and return to the
DATA
menu.
4) Press the
key to return to the measurement screen to begin measuring
reference lengths.
8.4 Auto Zero/Calibration
The auto calibration feature uses a single calibration bar and two multiples (echo-
echo mode) of the calibration bar to perform an automatic
Two-Point
calibration. If a
3” calibration bar is used, it would measure at 3”, and then again on the second
multiple at 6”. This zero option will typically be the preferred option, as it’s the easiest
option to use and for the most part is automatic in terms of operation. However,
depending on signal quality and calibration standard used, may not always work. If
valid signals cannot be acquired during the auto set routine, the
MiniMax
will report
“Can’t find stable echo”. Before eliminating this option altogether and using one of
the manual options listed below, try relocating the transducer to various positions on
the end of the calibration standard first. Since the calibration standards are
cylindrical, they can sometimes create substantial sidewall noise, based on the the
length, transducer size, and frequency. Moving the transducer to the edge of the
cylinder will often times eliminate the noise by shooting down the edge of the
standard and allowing the
MiniMax
to acquire a reasonable signal.
Now that the bolt group
CAL DATA
has been created in the previous section, we’re
ready to get started with the automatic zero/calibration process. The first thing to
note is the procedures in this section demonstrate a
Two-Point
calibration, but a very
fast and efficient one, with minimal adjustments required from the operator.
Measuring Calibration Standard – For the First Time