Alarms
and Actions
53
Item
Description
Indicators
Alarm sounds
Alarm screen appears in yellow on the
LCD display.
LCD state display: “K.V.O”
Alarm clearing
method
Press any buttons to clear the K.V.O alarm
If not cleared, the alarm sound will be heard every 15 seconds.
7.13
Complete Alarm
구분
설명
Cause
When the flow volume reaches the set amount, injection is stopped and
the standby screen appears. When K.V.O is set to ON, an injection K.V.O
alarm occurs.
Indicators
Alarm sounds
Alarm screen appears in yellow on the
LCD display.
LCD state display: “Complete”
Alarm clearing
method
Press any buttons to clear the K.V.O alarm
If not cleared, the alarm sound will be heard every 15 seconds.
7.14
Near complete alarm
Item
Description
Cause
When the injection completion time is within 5 minutes based on the total
injection volume and the volume to be injected, a near complete alarm
occurs.
You can set the duration time of the Near Complete Alarm.