Custom
Settings
40
No.
Name
Description
function.
Its setting is for single use only. So it must be set for each use.
▪ See
4.6 Infusion Relay Mode
Caution
Make sure that the I.V set is available. If inappropriate I.V set is used, an
occlusion or air alarm occurs or the specified flow rate may not be reached.
Mount the I.V Set without bending or twisting. Incorrect installation may keep
it from reaching the set rate.
Use the clamp to lock the I.V Set before opening the door or removing the I.V
set. Failure to close completely may cause water leakage.
Replace the drug container or I.V Set according to the maintenance procedure.
Install the drug container more than 50 cm higher than the patient's heart.
Do not use an I.V Set in the same location for more than 12 hours. Prolonged
use at the same location may cause deformation of the tube and injection
failure. It is recommended to move 15 cm every 12 hours or replace the set
with a new one.
6.3
Dosage Setting
The Dosage Setting includes the menus for F.Rate Calculation, Drug Library Use,
Frequently Used Drug and Drug Registration.
No.
Name
Description
1
F.Rate Calculation
Sets the infusion of the Dosage Infusion Mode.
▪ See
4.4 Dosage Infusion.
2
Drug Library Use
Selects the drug to set the infusion information of the Drug Inf
usion Mode
▪ See
4.3 Drug Infusion
.
3
Frequently Used Dr
ug
Registered drugs can be seen here functioning as bookmark.
4
Drug Registration
Selects a ward and search a drug to register for Frequently Us
ed Drug.