Alarms
and Actions
50
Item
Description
Alarm clearing
method
Press any buttons to clear the air bubble alarm.
Action
Make sure that the I.V Set is free of air when an air bubble alarm occurs
before use.
7.6
Door open alarm
Item
Description
Cause
When the door is open during injection, the door LED lights and the door
open alarm occurs.
Indicators
Alarm sounds
Alarm screen appears in red on the
LCD display.
LCD state display: “Door”
Alarm clearing
method
Press any buttons to clear the door open alarm.
Action
Make sure the door remains closed during injection.
7.7
Drop sensor alarm
Item
Description
Cause
The drop sensor alarm occurs in the following cases:
Unexpected free flow, leakage, empty container, etc. are detected
during injection.
During injection, significantly more or less drops are detected than the
specified value.
Indicators
Alarm sounds
Alarm screen appears in red on the
LCD display.
LCD state display: “Drop”
Alarm clearing
method
Press any buttons to clear the drop sensor alarm.
Action
Make sure the drop sensor is properly installed.