![Daiwha DI-6000 Скачать руководство пользователя страница 10](http://html1.mh-extra.com/html/daiwha/di-6000/di-6000_user-manual_3237401010.webp)
Safety
Instructions
4
Do not reuse I.V Set.
If I.V Set is used again, this may cause failures of sensors and infusion volume.
When using for a long time (more than 12 hours), change the location of I.V Set
Tube by 15 cm or more. Use it within the error range as instructed.
Do not mount I.V Set with extended.
Error may occur in flow rate or infusion volume.
Mount I.V Set in place.
Free flow or leakage may occur if not accurately mounted.
Do not connect other devices that can inject drugs in the I.V Set line (Patient Line).
It may affect the flow rate, infusion volume, and alarming.
1.4
Responsibilities of manufacturer
This product is a medical device. The manufacture or dealer has responsibility for normal
operation and safety in the following cases:
If genuine parts approved by the company
If serviced or repaired by the company or an agency delegated by the company
If properly used as instructed
If the product is damaged not by user fault
1.5
User obligations
The user must regularly maintain the product for electrical and mechanical safety (Common
electrical and mechanical standards (Ministry of Food and Drug Safety No. 2015-115) / IEC
60601- 1:2012). The owner of the product must guarantee the maintenance or repair work.
For more information, see 8. Maintenance and Storage.