![Daiwha DI-6000 Скачать руководство пользователя страница 14](http://html1.mh-extra.com/html/daiwha/di-6000/di-6000_user-manual_3237401014.webp)
Names
and Functions
8
2.3
Names and functions of each button
UI
Name
Function
JOG
Turn the
JOG
dial to choose a menu item or to adjust its value
Press the
JOG
button for saving the adjusted value of a menu
item.
JOG
only moves the cursor from the left to the right in digits. (Use
MUTE(right) button to move the cursor to the right)
Turn
JOG
clockwise to increase the digit number. Turn
JOG
counterclosewise to decrease the digit number.
BOLUS
/SAVE
When the
BOLUS
button is pressed during infusion or paused
state, infusion is performed by the flow rate and volume set at the
Infusion Setting Menu (BOLUS volume and flow rate).
To change the setting, press
BOLUS/SAVE
button.
To Save the adjusted setting, push
BOLUS/SAVE
or
JOG
to set
OK.
POWER
When the
POWER
button is pressed, the power of the product
immediately turns on.
The power of the product turns off by pressing and holding The
POWER
button.
CLEAR
/BACK
Cancels a setup operation or returns to the upper menu from a sub
menu when the
CLEAR/BACK
button is pressed.
Used to reset the input values in the setup menu.
START
/STOP
After the setup, if you press the
START/STOP
button, it starts
infusion.
If the
START/STOP
button is pressed again, infusion is paused.
MENU
Moves to the setup mode for the selected sub menu.
Moves the cursor to the left for setting the input values.
MUTE
Turns off the alarm when the alarm rings.
Moves the cursor to the right for setting the input values
LOCK
When
LOCK
button is pressed, the buttons other than
START/STOP
button and
POWER
button do not operate.