Custom
Settings
42
3
AC/DC Remove
Alarm
Used to set the alarm to On or Off when the AC/DC po
wer cable is connected or disconnected.
4
Alarm Volume
Used to set the alarm and sound volume.
Setting range is 1 ~ 10 steps. (50dB ~ 67dB)
5
Key Volume
Used to set the button sound volume.
6.4.2
Sensor
1
Up Occlusion level
Used to set the Up Stream Occlusion sensitivity level to
1 through 10.
Operation of ‘Up Occlusion’ may vary depending on the
quality of the ‘I.V Set’.
2
Down Occlusion le
vel
Used to set the Down Stream Occlusion sensitivity level
to 1 through 10.
3
Air Set
Used to set the air bubble size.
Air Set Level
Air Bubble Check
OFF
No detection
Large
15 mm or above
Medium
10 mm or above
Small
5 mm or above
4
Drop Sensitivity
Used to set the drop sensor’s sensitivity.
▪ You can set the sensitivity to High (90 % - sensitive),
Middle (100 % - default), or Low (110 % - insensitive).
6.4.3
History
1
Alarm log
Displays the alarm logs.
The list is sorted from the recently recorded items.
2
Event log
Displays the event logs.
The list is sorted from the recently recorded items.
6.4.4
Etc
1
Nurse call
Used to set the Nurse Call mode to On or Off when an
alarm occurs.
When the Nurse call is turned on, the pump will transmit
alarm signals to a nurse call system.
The nurse call system should be connected to the Nurse
Call connector on the back of the pump to alert nurses.
2
Auto Save
Used to set the Save function to On or Off