Alarms
and Actions
47
7.
Alarms and Actions
Alarms are triggered when the product is not functioning as intended. The
following describes required actions for each alarm. The system automatically
stops injection when an alarm occurs (according to the priority).
Be sure to check alarm messages, indication lamps, and alarm sounds to determine
required actions.
Caution
The operator must be in a location to monitor the product status while in
operation.
7.1
Alarm devices
There are various alarm devices to ensure patient safety. Depending on the alarm, these
devices send beep sounds, display alarm lamps, and notify the current status.
Alarm LCD
Red LCD ON:
High level of Severity
Yellow LCD ON
: Medium and Low level of Severity
You can check the product status within 1 meter through the state indicators and their alarms.
The situation indicators and their alarms can be checked within 3 meters.
The generated alarms can be cleared by pressing the STOP button. Once the relevant checklist
is filled in, press the START button to resume injection, and then check the alarm.
Caution
The product stops injection immediately upon detection of malfunctioning.