To determine a “time equivalent”, consult the laminated Time Chart that is provided and find the power 

output on the top row (horizontal axis). The numbers in the left most column (vertical axis) represent a 
range of potential treatment doses. Find the dose then trace that row to the right until it intersects with 
the column corresponding to the factory power output of the device. The time equivalent is shown where 
the  two  lines  intersect.  For  example,  if  the  factory  power  output  is  8.0  mW/cm

2

  and  the  dose  to  be 

delivered is 2 Joules, then the appropriate treatment time will be 04:10. A partial time chart illustrating 
this is shown on the following page.  

 

Note:

 For your convenience, and as an alternative to the UV time chart, the latest version of our free 

iPhone and Android application “Phototherapy Math” is available for download at the iTunes app store 

and Android market. It quickly calculates treatment times, backup safety times, and makes the 
calculation of dose and time increases simple. If using a PC, visit www.phototherapymath.com for a web 
based calculator and additional information.  
 

14.2

 

Partial UVA, UVA-1 Time Chart 

mW/cm2  

10 

12 

14 

Joules 

1.0 

03:20  02:47  02:23 

02:05 

01:51  01:40  01:23  01:11 

2.0 

06:40  05:33  04:46  04:10**  03:42  03:20  02:47  02:23 

3.0 

10:00  08:20  07:09 

06:15 

05:33  05:00  04:10  03:34 

 
The UVA and UVA-1 Time Chart (not shown here) is configured the same as the UVB Time Chart. The doses 
are listed in Joules and the energy output levels and times are different than the UVB output levels. Each 
chart is clearly marked. 
 

14.3

 

Setting Up a Treatment Time 

To administer a treatment you must enter your desired treatment time. Based on the example above the 
treatment time will be 4:10. 

1.

 

Press any key to awaken the unit. The word “CODE” will appear.

 

2.

 

Enter in your key code.  Unless you have changed it, the factory preset key code is 0007. 

3.

 

Once the key code has been entered, the display should read 00:00. 

4.

 

Press the 

 key until the first digit of your desired time is displayed. If the treatment time is one 

minute and ten seconds (04

:10), the first digit will be a “

4

”. 

 

5.

 

Press the 

 key 

to move the “

4

” one space to the left.  The display should read 00:

40. 

6.

 

Press the 

 

key until the next desired number is displayed (“1” in this case). The display should 

read 00:41. 

7.

 

Press the 

 key 

to move the “

4

” one space to the left. 

The display will now read 04:10. 

8.

 

When  the  desired  time  is  displayed  (in  this  case  04:10)  press  the  ENTER  key  to  lock  in  your 
treatment time. 

Put on your goggles and position yourself per the instructions.  

 

Содержание ML24000

Страница 1: ...MNL 00024 1 Proprietary and Confidential Page 1 of 21 OPERATION AND SERVICE MANUAL...

Страница 2: ...e Bodily Injury Hazards 11 11 3 Equipment Property Damage Hazards 11 12 0 Unlocking the Device 12 13 0 Changing or Disabling the Key Code 12 14 0 Flex Timer Instructions 12 14 1 Determining a Treatmen...

Страница 3: ...and Fitting Lamps 18 16 6 Resetting Your Lamp Hours 18 16 7 Troubleshooting 19 16 8 Spare Parts and Accessories 19 17 0 Warranty 20 17 1 Limited Warranty Policy 20 17 2 Warranty Coverage 20 17 3 Cust...

Страница 4: ...r Customer Service department and your input will be channeled to the appropriate person The purpose of this manual is to instruct users on the proper methods of operation and general maintenance In a...

Страница 5: ...Class II Mode of Operation Non continuous IEC EN 62471 2006 UV Risk Group Risk Group 2 Moderate Risk WARNING This device is designed for intermittent operation only and not for continuous use The devi...

Страница 6: ...into an outlet on a circuit different from that to which the other device s is connected Consult the manufacturer or field service technician for help 7 0 Electrical Specifications The device is manuf...

Страница 7: ...of your device as measured at the factory is shown on the Power Output Certificate included with your device literature The table below represents the possible output range of your unit Lamp Style Hi...

Страница 8: ...of manufacture that is specific to your device The ML24000 features a Data Connection port located on the Device This port is only to be used as a data only connection to the ML24000 Flex Controller...

Страница 9: ...and their definitions DANGEROUS VOLTAGE ATTENTION CONSULT ACCOMPANYING DOCUMENTS NON IONIZING RADIATION EARTH ground PROTECTIVE EARTH ground ALTERNATING CURRENT THREE PHASE ALTERNATING CURRENT WITH NE...

Страница 10: ...d licensed electrician must wire the service for this device in accordance with all national and local codes and the electrical instructions provided in the accompanying Service Installation Instructi...

Страница 11: ...t to avoid over exposure to isolated areas of the body 11 3 Equipment Property Damage Hazards Warning A qualified licensed electrician must wire the service for this device in accordance with all nati...

Страница 12: ...lf lock when left idle for twenty 20 minutes To unlock your unit follow the instructions below 1 Press any key to awaken your unit 2 Your device will display the word CODE 3 The factory preset key cod...

Страница 13: ...7 8 9 10 12 14 Joules 1 0 03 20 02 47 02 23 02 05 01 51 01 40 01 23 01 11 2 0 06 40 05 33 04 46 04 10 03 42 03 20 02 47 02 23 3 0 10 00 08 20 07 09 06 15 05 33 05 00 04 10 03 34 The UVA and UVA 1 Tim...

Страница 14: ...ake a mistake while entering a dose or backup time repeatedly press the key until all zeros are displayed To check the age in hours of your lamps press and release the MODE and keys together at the sa...

Страница 15: ...go more and more aggressive treatment just to overcome the natural UV screening effect of tanned skin This results in an excessively high cumulative dose to clear the patient s disease When setting up...

Страница 16: ...tightly across the skin Use two or more patches if more than six exposure levels are required Determine the doses you will test for recommended doses below Open exposure window 1 Begin the test by exp...

Страница 17: ...hese filters without proper training as certain solutions may damage the filters Only Daavlin authorized personnel should perform this function Allow the filters to cool down Use a vacuum to remove du...

Страница 18: ...rt the new lamps into the lamp sockets 10 Replace the filters in reverse order that they were removed 11 Replace the air filter brackets and acrylic for each column 12 Reconnect the unit at the breake...

Страница 19: ...eck all plug connections Check all fuses in the unit and the building If the lamp is defective change the lamp If the ignition unit is defective contact the Daavlin Service Department If it is impossi...

Страница 20: ...e misuse or accident incorrect customer installation improper electrical service lack of proper maintenance and Acts of God This Limited Warranty does not extend to repairs made necessary by use of pa...

Страница 21: ...rform any repair or maintenance on the Equipment unless prior written authorization has been obtained from Daavlin Even with Daavlin s authorization Daavlin shall not be responsible or liable for any...

Отзывы: