Begin the treatment by pressing ENTER.
To repeat the same dose for another exposure, press the END key to clear the display. Then press the
MODE and POWER keys at the same time and release. This will automatically bring up the last treatment
administered. Press ENTER to start the treatment.
Note:
The Digital Timer will beep three times to signal the end of a treatment. It will also flash the
treatment time on the display.
14.4
Special Functions of the Digital Timer
•
To pause an active treatment press either the END or ENTER key.
•
When a treatment is paused the screen will read “HOLD”. To resume a paused treatment press
the ENTER key.
•
To end or cancel an active treatment press the END key two times.
•
If you make a mistake while entering a dose or backup time, repeatedly press the
◄
key until all
zeros are displayed.
•
To check the age (in hours) of your lamps press and release the MODE and
◄
keys together at the
same time. The lamps must be off to do this.
14.5
Special Notes
•
The maximum time that can be entered is 59 minutes and 59 seconds
•
It is
not
possible to enter 60 seconds. You must enter 1:00 minute instead.
•
A short beep is sounded for every valid key entered. A long beep is sounded when an invalid key
is entered.
•
If power goes off during a treatment, the system will remember how much of the treatment has
elapsed. When power is restored, your device will be on “HOLD”. To continue the treatment,
press START. To cancel the treatment press END.
•
The small orange light above the display will be lit when a treatment is in process. It will blink
when the treatment is paused.
14.6
Flex Timer Range and Accuracy
Time Range:
1 Second - 59 Minutes 59 Seconds (00:01 - 59:59)
Accuracy:
+ / - 5%