5.0

 

Transport and Storage Specifications 

Ambient Temperature: 

-40°C to 70°C  (-40°F to 158°F) 

Relative Humidity: 

10% to 95%, Non-condensing  

Atmospheric  Pressure: 

50 kPa to 106 kPa 

Altitude: 

≤ 2000m

 

 

6.0

 

EMC Precautions 

The devices contained in this manual have been tested and found to comply with the EMC limits of the 
international standard IEC 60601-1-2.  These limits are designed to provide reasonable protection 
against interference in a typical medical installation.  The system can radiate radio frequency energy if 
not installed in accordance with the instructions, and may cause harmful interference to other devices in 
the vicinity.  There is no guarantee that interference will not occur in a particular installation.  If the 
system does cause interference with other devices, which can be determined by turning the system off 
and on, the user is encouraged to try to correct the interference by one or more of the following 
measures: 

 

Re-orient or relocate the receiving device 

 

Increase the separation between the equipment 

 

Connect the System into an outlet on a circuit different from that to which the other device(s) is 
connected 

Consult the manufacturer or field service technician for help 
 

7.0

 

Electrical Specifications 

The device is manufactured in the following electrical configurations.  Always refer to the identification 
label of the device to determine which electrical configuration the device is manufactured to. 
 

 

Mains input = 230V, 3~, 60Hz, 100A   Operating = 230V, 3~, phase to phase  

 

Mains input = 400V, 3N~, 50Hz, 100A   Operating = 230V~, single-phase (phase to neutral) 

 

8.0

 

Site Selection 

A site should be chosen within reach of the specified electrical connection (refer to the Facility 
Requirements Guide) and where the unit can be left in place without obstructing traffic flow. It is 
important that the unit be properly grounded. The site should not be in any area where water or 
moisture might collect and should be protected from access by children and other unintended users.  

WARNING: 

To avoid the risk of electric shock this equipment must only be connected to a supply mains 

with a protective earth.  

CAUTION:

 This device is a Class A Medical Device suitable for use in all establishments other than 

domestic and those directly connected to the public low voltage power supply network that supplies 
buildings used for domestic purposes. 

Содержание ML24000

Страница 1: ...MNL 00024 1 Proprietary and Confidential Page 1 of 21 OPERATION AND SERVICE MANUAL...

Страница 2: ...e Bodily Injury Hazards 11 11 3 Equipment Property Damage Hazards 11 12 0 Unlocking the Device 12 13 0 Changing or Disabling the Key Code 12 14 0 Flex Timer Instructions 12 14 1 Determining a Treatmen...

Страница 3: ...and Fitting Lamps 18 16 6 Resetting Your Lamp Hours 18 16 7 Troubleshooting 19 16 8 Spare Parts and Accessories 19 17 0 Warranty 20 17 1 Limited Warranty Policy 20 17 2 Warranty Coverage 20 17 3 Cust...

Страница 4: ...r Customer Service department and your input will be channeled to the appropriate person The purpose of this manual is to instruct users on the proper methods of operation and general maintenance In a...

Страница 5: ...Class II Mode of Operation Non continuous IEC EN 62471 2006 UV Risk Group Risk Group 2 Moderate Risk WARNING This device is designed for intermittent operation only and not for continuous use The devi...

Страница 6: ...into an outlet on a circuit different from that to which the other device s is connected Consult the manufacturer or field service technician for help 7 0 Electrical Specifications The device is manuf...

Страница 7: ...of your device as measured at the factory is shown on the Power Output Certificate included with your device literature The table below represents the possible output range of your unit Lamp Style Hi...

Страница 8: ...of manufacture that is specific to your device The ML24000 features a Data Connection port located on the Device This port is only to be used as a data only connection to the ML24000 Flex Controller...

Страница 9: ...and their definitions DANGEROUS VOLTAGE ATTENTION CONSULT ACCOMPANYING DOCUMENTS NON IONIZING RADIATION EARTH ground PROTECTIVE EARTH ground ALTERNATING CURRENT THREE PHASE ALTERNATING CURRENT WITH NE...

Страница 10: ...d licensed electrician must wire the service for this device in accordance with all national and local codes and the electrical instructions provided in the accompanying Service Installation Instructi...

Страница 11: ...t to avoid over exposure to isolated areas of the body 11 3 Equipment Property Damage Hazards Warning A qualified licensed electrician must wire the service for this device in accordance with all nati...

Страница 12: ...lf lock when left idle for twenty 20 minutes To unlock your unit follow the instructions below 1 Press any key to awaken your unit 2 Your device will display the word CODE 3 The factory preset key cod...

Страница 13: ...7 8 9 10 12 14 Joules 1 0 03 20 02 47 02 23 02 05 01 51 01 40 01 23 01 11 2 0 06 40 05 33 04 46 04 10 03 42 03 20 02 47 02 23 3 0 10 00 08 20 07 09 06 15 05 33 05 00 04 10 03 34 The UVA and UVA 1 Tim...

Страница 14: ...ake a mistake while entering a dose or backup time repeatedly press the key until all zeros are displayed To check the age in hours of your lamps press and release the MODE and keys together at the sa...

Страница 15: ...go more and more aggressive treatment just to overcome the natural UV screening effect of tanned skin This results in an excessively high cumulative dose to clear the patient s disease When setting up...

Страница 16: ...tightly across the skin Use two or more patches if more than six exposure levels are required Determine the doses you will test for recommended doses below Open exposure window 1 Begin the test by exp...

Страница 17: ...hese filters without proper training as certain solutions may damage the filters Only Daavlin authorized personnel should perform this function Allow the filters to cool down Use a vacuum to remove du...

Страница 18: ...rt the new lamps into the lamp sockets 10 Replace the filters in reverse order that they were removed 11 Replace the air filter brackets and acrylic for each column 12 Reconnect the unit at the breake...

Страница 19: ...eck all plug connections Check all fuses in the unit and the building If the lamp is defective change the lamp If the ignition unit is defective contact the Daavlin Service Department If it is impossi...

Страница 20: ...e misuse or accident incorrect customer installation improper electrical service lack of proper maintenance and Acts of God This Limited Warranty does not extend to repairs made necessary by use of pa...

Страница 21: ...rform any repair or maintenance on the Equipment unless prior written authorization has been obtained from Daavlin Even with Daavlin s authorization Daavlin shall not be responsible or liable for any...

Отзывы: