MNL-00024 [1] 

Proprietary and Confidential  

Page 

15

 of 

21 

 

15.0

 

Treatment Protocols 

15.1

 

MED Testing 

The more accurate but more complicated method of determining a patient's initial dose is through MED 
testing. Because this is a cumbersome process, many doctors do not wish to determine a patient's initial 
dose in this manner. However, before we discuss the procedure in greater detail, let us first define the 
terms. 
 
MED  stands  for  Minimum  Erythemal  Dose  and  is  defined  as  the  minimal  dose  of  UVB  that  will  cause 
marked erythema on a patient's skin. As an example, a skin type IV will have an MED somewhere within 
the range of 200-600 milliJoules of narrow band UVB. See Protocols for all skin types and ranges. 
 
MED testing  results  in  a  more  accurate  starting  point.  Because of  the  subjective  nature  of  a  patient's 
response to the skin-typing quiz, the empirical method is more subject to error. In many cases, testing will 
show that a patient may be more tolerant to ultraviolet than would have been suggested by simple skin 
typing. 
 
It is important to start phototherapy patients at as high a dose as possible. Underdosing a patient causes 
excessive tanning that means that the patient must undergo more and more aggressive treatment just to 
overcome the natural UV screening effect of tanned skin. This results in an excessively high cumulative 
dose to clear the patient's disease. 
 
When setting up a UVA-1 patient, the key element in determining exactly how much UVA-1 radiation the 
patient will receive is the MED. The first UVA-1 treatment is always a percentage of the MED that has 
been entered. Thereafter, each succeeding treatment is some percentage of the preceding dose. Each and 
every dose is mathematically related to the MED value, whether this value was determined by MED testing 
or by skin typing. 
 
The following section provides instructions on exactly how MED testing is carried out in the doctor's office. 
 

15.2

 

MED Testing Procedure 

The 

first stage in performing MED testing is to determine the patient’s skin type.  The patient’s skin type 

is then used to determine the dosages that will be used to perform MED testing.  Many physicians use a 
simple table of questions to determine a patient's tolerance to light. The table is reproduced below: 
 

Skin Type  

History 

Skin Type I  

Always burns, never tans 

Skin Type II  

Always burns, sometimes tans 

Skin Type III  

Sometimes burns, always tans 

Skin Type IV  

Always tans, never burns 

Skin Type V  

Brown skinned (Chinese, Mexican American) 

Skin Type VI  

Black 

 

Once you have determined the patient’s skin type you can proceed to the testing stage.

 

 

Содержание ML24000

Страница 1: ...MNL 00024 1 Proprietary and Confidential Page 1 of 21 OPERATION AND SERVICE MANUAL...

Страница 2: ...e Bodily Injury Hazards 11 11 3 Equipment Property Damage Hazards 11 12 0 Unlocking the Device 12 13 0 Changing or Disabling the Key Code 12 14 0 Flex Timer Instructions 12 14 1 Determining a Treatmen...

Страница 3: ...and Fitting Lamps 18 16 6 Resetting Your Lamp Hours 18 16 7 Troubleshooting 19 16 8 Spare Parts and Accessories 19 17 0 Warranty 20 17 1 Limited Warranty Policy 20 17 2 Warranty Coverage 20 17 3 Cust...

Страница 4: ...r Customer Service department and your input will be channeled to the appropriate person The purpose of this manual is to instruct users on the proper methods of operation and general maintenance In a...

Страница 5: ...Class II Mode of Operation Non continuous IEC EN 62471 2006 UV Risk Group Risk Group 2 Moderate Risk WARNING This device is designed for intermittent operation only and not for continuous use The devi...

Страница 6: ...into an outlet on a circuit different from that to which the other device s is connected Consult the manufacturer or field service technician for help 7 0 Electrical Specifications The device is manuf...

Страница 7: ...of your device as measured at the factory is shown on the Power Output Certificate included with your device literature The table below represents the possible output range of your unit Lamp Style Hi...

Страница 8: ...of manufacture that is specific to your device The ML24000 features a Data Connection port located on the Device This port is only to be used as a data only connection to the ML24000 Flex Controller...

Страница 9: ...and their definitions DANGEROUS VOLTAGE ATTENTION CONSULT ACCOMPANYING DOCUMENTS NON IONIZING RADIATION EARTH ground PROTECTIVE EARTH ground ALTERNATING CURRENT THREE PHASE ALTERNATING CURRENT WITH NE...

Страница 10: ...d licensed electrician must wire the service for this device in accordance with all national and local codes and the electrical instructions provided in the accompanying Service Installation Instructi...

Страница 11: ...t to avoid over exposure to isolated areas of the body 11 3 Equipment Property Damage Hazards Warning A qualified licensed electrician must wire the service for this device in accordance with all nati...

Страница 12: ...lf lock when left idle for twenty 20 minutes To unlock your unit follow the instructions below 1 Press any key to awaken your unit 2 Your device will display the word CODE 3 The factory preset key cod...

Страница 13: ...7 8 9 10 12 14 Joules 1 0 03 20 02 47 02 23 02 05 01 51 01 40 01 23 01 11 2 0 06 40 05 33 04 46 04 10 03 42 03 20 02 47 02 23 3 0 10 00 08 20 07 09 06 15 05 33 05 00 04 10 03 34 The UVA and UVA 1 Tim...

Страница 14: ...ake a mistake while entering a dose or backup time repeatedly press the key until all zeros are displayed To check the age in hours of your lamps press and release the MODE and keys together at the sa...

Страница 15: ...go more and more aggressive treatment just to overcome the natural UV screening effect of tanned skin This results in an excessively high cumulative dose to clear the patient s disease When setting up...

Страница 16: ...tightly across the skin Use two or more patches if more than six exposure levels are required Determine the doses you will test for recommended doses below Open exposure window 1 Begin the test by exp...

Страница 17: ...hese filters without proper training as certain solutions may damage the filters Only Daavlin authorized personnel should perform this function Allow the filters to cool down Use a vacuum to remove du...

Страница 18: ...rt the new lamps into the lamp sockets 10 Replace the filters in reverse order that they were removed 11 Replace the air filter brackets and acrylic for each column 12 Reconnect the unit at the breake...

Страница 19: ...eck all plug connections Check all fuses in the unit and the building If the lamp is defective change the lamp If the ignition unit is defective contact the Daavlin Service Department If it is impossi...

Страница 20: ...e misuse or accident incorrect customer installation improper electrical service lack of proper maintenance and Acts of God This Limited Warranty does not extend to repairs made necessary by use of pa...

Страница 21: ...rform any repair or maintenance on the Equipment unless prior written authorization has been obtained from Daavlin Even with Daavlin s authorization Daavlin shall not be responsible or liable for any...

Отзывы: