16.5
Removing and Fitting Lamps
1.
Measure lamp intensity of current lamp before replacement.
2.
Switch the lamp off; allow the lamps to cool down.
3.
Turn unit off at the breaker.
4.
Remove the air filter brackets and acrylic from each column.
5.
Put on latex gloves when handling filters and lamps.
6.
Remove the three clear filters from each column and place in safe location.
7.
Remove the six colored filters from each column and place in safe location.
8.
Grasp the end of the lamp and pull from lamp socket.
9.
Insert the new lamps into the lamp sockets.
10.
Replace the filters in reverse order that they were removed.
11.
Replace the air filter brackets and acrylic for each column.
12.
Reconnect the unit at the breaker.
13.
Switch on the unit and measure UV intensity in order to be able to redefine the necessary
exposure times.
Notify all those involved with the unit’s operation of the lamp replacement as patient exposure times
may need adjusting.
16.6
Resetting Your Lamp Hours
Whenever the lamps are changed in your device, it is important to reset the lamp age to zero.
To Reset lamp hours to zero when installing new lamps, take the following steps:
Press MODE, then
◄
and ENTER all at the same time
Then…
Press
▲
Then…
Press ENTER
Then…
Press MODE and
◄
at the same time
Then…
Press END