background image

MNL-00024 [1] 

Proprietary and Confidential  

Page 

7

 of 

21 

 

9.0

 

UVA-1 Output Spectrum 

The treatment light is generated by 24 1000 Watt High Intensity Discharge Metal Halide Lamps. The light 
emitted is in the UVA band with a primary emission range of 350-400 nanometers. The output spectrum 
of the system is shown in Figure 1.4. Some of the light outside the 350-400nm peak emission range is 
typical for these types of systems and does not affect safety or efficiency. 
 

 

Figure 1 

9.1

 

Irradiance Specifications 

The power output of your unit is dependent on the style and quantity of lamps contained within the 
device. The power output of your unit is fixed and cannot be altered by the user. The power output 
value of your device, as measured at the factory, is shown on the 

Power Output Certificate

 included with 

your device literature. The table below represents the possible output range of your unit. 
 

Lamp Style: 

High Intensity Discharge 

Minimum 

Output: 

9" Away: 60 mW/cm

2

 (+/- 10%)  

Maximum 

Output: 

9" Away: 100 mW/cm

2

 (+/- 10%)  

 

Note: 

Irradiance measurements taken with Gigahertz-Optik P-9710-1 Optometer. 

 

9.2

 

Labels and Symbols 

A warning label is affixed to your device in a prominent and easily readable position. Please read the 
label carefully as it contains important safety information for you. According to IEC/EN 62471:2006, 
sources of optical radiation are classified into risk groups subject to their potential photobiological 
hazard. This device falls under Risk Group 2 (Moderate-Risk) and does not pose a severe hazard due to 
the natural aversion response to bright light or thermal discomfort. A Risk Group 2 warning label, which 

Содержание ML24000

Страница 1: ...MNL 00024 1 Proprietary and Confidential Page 1 of 21 OPERATION AND SERVICE MANUAL...

Страница 2: ...e Bodily Injury Hazards 11 11 3 Equipment Property Damage Hazards 11 12 0 Unlocking the Device 12 13 0 Changing or Disabling the Key Code 12 14 0 Flex Timer Instructions 12 14 1 Determining a Treatmen...

Страница 3: ...and Fitting Lamps 18 16 6 Resetting Your Lamp Hours 18 16 7 Troubleshooting 19 16 8 Spare Parts and Accessories 19 17 0 Warranty 20 17 1 Limited Warranty Policy 20 17 2 Warranty Coverage 20 17 3 Cust...

Страница 4: ...r Customer Service department and your input will be channeled to the appropriate person The purpose of this manual is to instruct users on the proper methods of operation and general maintenance In a...

Страница 5: ...Class II Mode of Operation Non continuous IEC EN 62471 2006 UV Risk Group Risk Group 2 Moderate Risk WARNING This device is designed for intermittent operation only and not for continuous use The devi...

Страница 6: ...into an outlet on a circuit different from that to which the other device s is connected Consult the manufacturer or field service technician for help 7 0 Electrical Specifications The device is manuf...

Страница 7: ...of your device as measured at the factory is shown on the Power Output Certificate included with your device literature The table below represents the possible output range of your unit Lamp Style Hi...

Страница 8: ...of manufacture that is specific to your device The ML24000 features a Data Connection port located on the Device This port is only to be used as a data only connection to the ML24000 Flex Controller...

Страница 9: ...and their definitions DANGEROUS VOLTAGE ATTENTION CONSULT ACCOMPANYING DOCUMENTS NON IONIZING RADIATION EARTH ground PROTECTIVE EARTH ground ALTERNATING CURRENT THREE PHASE ALTERNATING CURRENT WITH NE...

Страница 10: ...d licensed electrician must wire the service for this device in accordance with all national and local codes and the electrical instructions provided in the accompanying Service Installation Instructi...

Страница 11: ...t to avoid over exposure to isolated areas of the body 11 3 Equipment Property Damage Hazards Warning A qualified licensed electrician must wire the service for this device in accordance with all nati...

Страница 12: ...lf lock when left idle for twenty 20 minutes To unlock your unit follow the instructions below 1 Press any key to awaken your unit 2 Your device will display the word CODE 3 The factory preset key cod...

Страница 13: ...7 8 9 10 12 14 Joules 1 0 03 20 02 47 02 23 02 05 01 51 01 40 01 23 01 11 2 0 06 40 05 33 04 46 04 10 03 42 03 20 02 47 02 23 3 0 10 00 08 20 07 09 06 15 05 33 05 00 04 10 03 34 The UVA and UVA 1 Tim...

Страница 14: ...ake a mistake while entering a dose or backup time repeatedly press the key until all zeros are displayed To check the age in hours of your lamps press and release the MODE and keys together at the sa...

Страница 15: ...go more and more aggressive treatment just to overcome the natural UV screening effect of tanned skin This results in an excessively high cumulative dose to clear the patient s disease When setting up...

Страница 16: ...tightly across the skin Use two or more patches if more than six exposure levels are required Determine the doses you will test for recommended doses below Open exposure window 1 Begin the test by exp...

Страница 17: ...hese filters without proper training as certain solutions may damage the filters Only Daavlin authorized personnel should perform this function Allow the filters to cool down Use a vacuum to remove du...

Страница 18: ...rt the new lamps into the lamp sockets 10 Replace the filters in reverse order that they were removed 11 Replace the air filter brackets and acrylic for each column 12 Reconnect the unit at the breake...

Страница 19: ...eck all plug connections Check all fuses in the unit and the building If the lamp is defective change the lamp If the ignition unit is defective contact the Daavlin Service Department If it is impossi...

Страница 20: ...e misuse or accident incorrect customer installation improper electrical service lack of proper maintenance and Acts of God This Limited Warranty does not extend to repairs made necessary by use of pa...

Страница 21: ...rform any repair or maintenance on the Equipment unless prior written authorization has been obtained from Daavlin Even with Daavlin s authorization Daavlin shall not be responsible or liable for any...