Caution: This device is a Class A Medical Device suitable for use in all establishments other than
domestic and those directly connected to the public low voltage power supply network that
supplies buildings used for domestic purposes.
•
Caution: This device is designed for intermittent operation only and not for continuous use. The
device should not be cycled continuously. For treatments greater than 20 minutes in duration
the device should be either turned off or left idling for a minimum of 50% of the administered
treatment time, e.g. a treatment lasting 40 minutes in duration should be followed by a cool
down period of 20 minutes.
12.0
Unlocking the Device
To prevent unauthorized use, the device will “self lock” when
left idle for twenty (20) minutes. To unlock
your unit follow the instructions below.
1.
Press any key to awaken your unit.
2.
Your device will display the word “CODE”.
3.
The factory preset key code is “0007”. Press the
▲
key seven times. Your unit will now display
“0007”.
4.
Press the “ENTER” key to unlock your unit.
13.0
Changing or Disabling the Key Code
If you would like to disable or personalize your factory default key code please contact the Daavlin
service department at 1-800-322-8546 for step by step instructions.
14.0
Flex Timer Instructions
14.1
Determining a Treatment Time
When using time as a dose, there are two treatment methods that can be used. The simplest is to start
at a prescribed time and then increase treatment times by a specific time increment or by a percentage.
Another method is to use the factory power output levels (shown on the
Power Output Certificate
) to
determine estimated “time equivalents” needed to deliver doses
in Joules (UVA-1).