52
Culligan® Series M2 Reverse Osmosis
52
Cat. No. 01023095
Screen Display
Range
Procedure
PS 1 = OPEN
PTLO = OPEN
FS HI = OPEN
FS LO = OPEN
This screen shows the switch input status for the Master unit/ First
Pass. PS1 represents the feed pressure switch. PTLO represents the
pre treat locout. FS HI represents the high level float switch. FS LOW
represents the low level float switch.
TOTAL FEED WATER
0
GALLONS
This screen shows the total feed water measured by feed flow meter
of the Master unit/ First Pass, since the the unit was put into service.
TOTAL PROD WATER
0
GALLONS
This screen shows the total product water of Master unit/ First Pass,
which was recorded by the Master unit/ First Pass product flow me-
ter, since the unit was put into service
PUMPING TIME
0
HOURS
This screen displays the total number of hours the Master unit/
First Pass has been in service (running), since the unit was put into
service.
ERROR: 1
Water Press Low
JAN-01-12 12:01P
This screen displays the last one to five errors flagged by the unit. If
no errors are flagged the display reads NONE.
FWR202LT04
Feb 1 2013
S/N:00003790
This screen shows the firmware version, born-on-date and serial
number for the Master unit.
NOTE Both units should have the SAME firmware version
and born-on-date.
HISTORICAL DATA
JAN 01
REJ=97% REC=50
DP = 77 PSI
100-1000
The GROC has a built in data logging feature which takes a “snap-
shot at the start of each month. The first time each month that the
system has been in a “RUNNING” state for a continuous 10 minute
block of time, the system will save the current values of:
Totalized gallons since new
Current % recovery
Current % rejection
Current Product TDS
Current water temperature
This data is stored on the local GROC board. The “snapshots”
from the first three months are always stored with the most recent
24-month “snapshots”. This historical data is available even if the
modem option is not used. If the historical data has been cleared, the
screen displays NONE.